Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2017 May;475(5):1472-1482.
doi: 10.1007/s11999-016-5182-8. Epub 2016 Nov 28.

What Are the Biomechanical Properties of the Taylor Spatial Frame™?

Affiliations
Comparative Study

What Are the Biomechanical Properties of the Taylor Spatial Frame™?

Daniel J Henderson et al. Clin Orthop Relat Res. 2017 May.

Abstract

Background: The Taylor Spatial Frame™ (TSF) is a versatile variant of the traditional Ilizarov circular fixator. Although in widespread use, little comparative data exist to quantify the biomechanical effect of substituting the tried-and-tested Ilizarov construct for the TSF hexapod system.

Questions/purposes: This study was designed to investigate the mechanical properties of the TSF system under physiologic loads, with and without the addition of a simulated bone model, with comparison to the standard Ilizarov frame.

Methods: The mechanical behaviors of three identical four-ring TSF and Ilizarov constructs were tested under levels of axial compression, bending, and rotational torque to simulate loading during normal gait. An acrylic-pipe fracture model subsequently was mounted, using fine wires and 5 mm half pins, and the testing was repeated. Load-deformation curves, and so rigidity, for each construct were calculated, with statistical comparisons performed using paired t-tests.

Results: Under axial loading, the TSF was found to be less rigid than the Ilizarov frame (645 ± 57 N/mm versus 1269 ± 256 N/mm; mean difference, 623 N/mm; 95% CI, 438.3-808.5 N/mm; p < 0.001), but more rigid under bending and torsional loads (bending: 42 ± 9 Nm/degree versus 78 ± 13 Nm/degree; mean difference, 37 Nm/degree; 95% CI, 25.0-47.9 Nm/degree; p < 0.001; torsion: 16 ± 2 Nm/degree versus 5 ± 0.35 Nm/degree; mean difference, 11 Nm/degree; 95% CI, 9.5-12.2 Nm/degree; p < 0.001). On mounting the bone models, these relationships broadly remained in the half-pin and fine-wire groups, however the half-pin constructs were universally more rigid than those using fine wires. This effect resulted in the TSF, using half pins, showing no difference in axial rigidity to the fine-wire Ilizarov (107 ± 3 N/mm versus 107 ± 4 N/mm; mean difference, 0.05 N/mm; 95% CI, -6.99 to 7.1 N/mm; p > 0.999), while retaining greater bending and torsional rigidity. Throughout testing, a small amount of laxity was observed in the TSF construct on either side of neutral loading, amounting to 0.72 mm (±0.37 mm) for a change in loading between -10 N and 10 N axial load, and which persisted with the addition of the synthetic fracture model.

Conclusions: This study broadly shows the TSF construct to generate lower axial rigidity, but greater bending and torsional rigidity, when compared with the Ilizarov frame, under physiologic loads. The anecdotally described laxity in the TSF hexapod strut system was shown in vitro, but only at low levels of loading around neutral. It also was shown that the increased stiffness generated by use of half pins produced a TSF construct replicating the axial rigidity of a fine-wire Ilizarov frame, for which much evidence of good clinical and radiologic outcomes exist, while providing greater rigidity and so improved resistance to potentially detrimental bending and rotational shear loads.

Clinical relevance: If replicated in the clinical setting, these findings suggest that when using the TSF, care should be taken to minimize the observed laxity around neutral with appropriate preloading of the construct, but that its use may produce constructs better able to resist bending and torsional loading, although with lower axial rigidity. Use of half pins in a TSF construct however may replicate the axial mechanical behavior of an Ilizarov construct, which is thought to be conducive to bone healing.

PubMed Disclaimer

Figures

Fig. 1A–F
Fig. 1A–F
The six frame constructs tested were the (A) Ilizarov alone, (B) Taylor Spatial Frame™ (TSF) alone, (C) fine-wire Ilizarov, (D) fine-wire TSF, (E) half-pin Ilizarov, and (F) half-pin TSF.
Fig. 2
Fig. 2
A fine-wire Taylor Spatial Frame™ mounted in the axial testing apparatus (Tinius Olsen H25K-S UTM; Tinius Olsen Inc) is shown.
Fig. 3A–C
Fig. 3A–C
Load deformation plots for testing of frame elements alone under (A) axial, (B) bending, and (C) torsional loads are shown. TSF = Taylor Spatial Frame™.
Fig. 4A–C
Fig. 4A–C
Box and whisker plots of stiffness data for Ilizarov and TSF elements alone, fine-wire, and half-pin constructs under (A) axial, (B) bending, and (C) torsional loading are shown. The boxes represent the 25th to 75th percentiles with whiskers from minimum to maximum values. The line through the middle of the box represents the median and + = the mean. TSF = Taylor Spatial Frame™.
Fig. 5A–C
Fig. 5A–C
Load deformation plots for testing of bone-frame constructs under (A) axial, (B) bending, and (C) torsional loads are shown. TSF = Taylor Spatial Frame™.

Comment in

References

    1. Al-Sayyad MJ. Taylor Spatial Frame in the treatment of pediatric and adolescent tibial shaft fractures. J Pediatr Orthop. 2006;26:164–170. doi: 10.1097/01.bpo.0000218522.05868.f9. - DOI - PubMed
    1. Antoci V, Voor MJ, Antoci V, Jr, Roberts CS. Biomechanics of olive wire positioning and tensioning characteristics. J Pediatr Orthop. 2005;25:798–803. doi: 10.1097/01.bpo.0000184646.03981.ec. - DOI - PubMed
    1. Augat P, Burger J, Schorlemmer S, Henke T, Peraus M, Claes L. Shear movement at the fracture site delays healing in a diaphyseal fracture model. J Orthop Res. 2003;21:1011–1017. doi: 10.1016/S0736-0266(03)00098-6. - DOI - PubMed
    1. Binski JC. Taylor Spatial Frame in acute fracture care. Tech Orthop. 2002;17:173–184. doi: 10.1097/00013611-200206000-00007. - DOI
    1. Board TN, Yang L, Saleh M. Why fine-wire fixators work: an analysis of pressure distribution at the wire-bone interface. J Biomech. 2007;40:20–25. doi: 10.1016/j.jbiomech.2005.12.005. - DOI - PubMed

Publication types