Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2016 Dec;43(12):6491.
doi: 10.1118/1.4967344.

Examining credentialing criteria and poor performance indicators for IROC Houston's anthropomorphic head and neck phantom

Affiliations

Examining credentialing criteria and poor performance indicators for IROC Houston's anthropomorphic head and neck phantom

Mallory E Carson et al. Med Phys. 2016 Dec.

Abstract

Purpose: To analyze the most recent results of the Imaging and Radiation Oncology Core Houston Quality Assurance Center's (IROC-H) anthropomorphic head and neck (H&N) phantom to determine the nature of failing irradiations and the feasibility of altering credentialing criteria.

Methods: IROC-H's H&N phantom, used for intensity-modulated radiation therapy credentialing for National Cancer Institute-sponsored clinical trials, requires that an institution's treatment plan agrees within ±7% of measured thermoluminescent dosimeter (TLD) doses; it also requires that ≥85% of pixels pass ±4 mm distance to agreement (7%/4 mm gamma analysis for film). The authors re-evaluated 156 phantom irradiations (November 1, 2014-October 31, 2015) according to the following tighter criteria: (1) 5% TLD and 5%/4 mm, (2) 5% TLD and 5%/3 mm, (3) 4% TLD and 4%/4 mm, and (4) 3% TLD and 3%/3 mm. Failure rates were evaluated with respect to individual film and TLD performance by location in the phantom. Overall poor phantom results were characterized qualitatively as systematic errors (correct shape and position but wrong magnitude of dose), setup errors/positional shifts, global but nonsystematic errors, and errors affecting only a local region.

Results: The pass rate for these phantoms using current criteria was 90%. Substituting criteria 1-4 reduced the overall pass rate to 77%, 70%, 63%, and 37%, respectively. Statistical analyses indicated that the probability of noise-induced TLD failure, even at the 5% criterion, was <0.5%. Phantom failures were generally identified by TLD (≥66% failed TLD, whereas ≥55% failed film), with most failures occurring in the primary planning target volume (≥77% of cases). Results failing current criteria or criteria 1 were primarily diagnosed as systematic >58% of the time (11/16 and 21/36 cases, respectively), with a greater extent due to underdosing. Setup/positioning errors were seen in 11%-13% of all failing cases (2/16 and 4/36 cases, respectively). Local errors (8/36 cases) could only be demonstrated at criteria 1. Only three cases of global errors were identified in these analyses. For current criteria and criteria 1, irradiations that failed from film only were overwhelmingly associated with phantom shifts/setup errors (≥80% of cases).

Conclusions: This study highlighted that the majority of phantom failures are the result of systematic dosimetric discrepancies between the treatment planning system and the delivered dose. Further work is necessary to diagnose and resolve such dosimetric inaccuracy. In addition, the authors found that 5% TLD and 5%/4 mm gamma criteria may be both practically and theoretically achievable as an alternative to current criteria.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
IROC-H H&N phantom with block dosimetry insert.
FIG. 2.
FIG. 2.
Sample global error. These plots of (a) a right–left profile taken from an axial film and (b) a superior–inferior profile taken from a sagittal film are from an institution trial deemed to exhibit global nonsystematic errors in treatment planning and delivery. This plan has poor dose uniformity through both the primary PTV and the secondary PTV. No additional error patterns (rotation, shifting, etc.) were evidenced in this irradiation. PTV = planning target volume.
FIG. 3.
FIG. 3.
Sample systematic error. These plots of (a) a superior–inferior profile taken from a sagittal film and (b) an anterior–posterior profile taken from an axial film are from an institution trial deemed to exhibit systematic errors in treatment planning and delivery, as there is a systematic difference of ∼8% between the measured and calculated dose distributions in the high-dose region. Such errors constitute the majority of errors observed in failing plans. PTV = planning target volume.

References

    1. Ibbott G. S., Followill D. S., Molineu H. A., Lowenstein J. R., Alvarez P. E., and Roll J. E., “Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials,” Int. J. Radiat. Oncol., Biol., Phys. (1), S71–S75 (2008).10.1016/j.ijrobp.2007.08.083 - DOI - PMC - PubMed
    1. Molineu A., Hernandez N., Nguyen T., Ibbott G., and Followill D., “Credentialing results from IMRT irradiations of an anthropomorphic head and neck phantom,” Med. Phys. (2), 022101 (2pp.) (2013).10.1118/1.4773309 - DOI - PMC - PubMed
    1. McVicker D., Yin F.-F., and Adamson J. D., “On the sensitivity of TG-119 and IROC credentialing to TPS commissioning errors,” J. Appl. Clin. Med. Phys. (1), 34–48 (2016).10.1120/jacmp.v17i1.5452 - DOI - PMC - PubMed
    1. Molineu A., Followill D. S., Balter P. A., Hanson W. F., Gillin M. T., Huq M. S., Eisbruch A., and Ibbott G. S., “Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the Radiation Therapy Oncology Group,” Int. J. Radiat. Oncol. (2), 577–583 (2005).10.1016/j.ijrobp.2005.05.021 - DOI - PubMed
    1. Ibbott G. S., Molineu A., and Followill D. S., “Independent evaluations of IMRT through the use of an anthropomorphic phantom,” Technol. Cancer Res. Treat. (5), 481–487 (2006).10.1177/153303460600500504 - DOI - PubMed

LinkOut - more resources