Measurement of heated skin O2 diffusion conductance and PO2 sensor induced O2 gradient
- PMID: 279189
Measurement of heated skin O2 diffusion conductance and PO2 sensor induced O2 gradient
Abstract
A large cathode heated transcutaneous PO2 electrode was used to measure the resistance of skin to diffusion of O2 from the capillary bed to the surface. With a non-O2-consuming tcPO2 electrode, one would read ideal tcPO2, Pi, which approximates PaO2 between 50 and 100 mmHg due to approximate cancellation of the rise by heating and the fall by skin O2 consumption. When a large cathode tcPO2 electrode is used, its O2 consumption causes a PO2 gradient across skin resistance, Rs, which depends on electrode membrane resistance RM (with Mylar), or RT (with Telfon). Surface PO2 will fall below Pi, and assuming no change in capillary PO2: PT/Pi = RT/(RT + Rs) and PM/Pi = RM/(RM + Rs) (1) 12 micrometer Telfon is 50 times as permeable to O2 as 6 micrometer Myler. When an electrode, covered alternately with these two membranes, was mounted on the same skin site in six subjects, the mean observed PT/PM was about 0.3. Substituting in equation (1) yields RT/Rs = 0.4, Rm/Rs = 20, and PM/Pi = 0.95. Using the known O2 permeability of Teflon, we computed skin O2 conductance to be 6.6 pmol O2/(cm2 .s .kPa), or 15 nl O2/(cm2 .s .atm). This method thus provides a way of determining ideal tc PO2 and stirring effect, phi, of other electrodes and various membranes.
Similar articles
-
Problems of calibration and stabilization of tcPO2 electrodes.Acta Anaesthesiol Scand Suppl. 1978;68:68-72. Acta Anaesthesiol Scand Suppl. 1978. PMID: 279203
-
Blood flow rate, temperature, oxygen tension and consumption in the skin of adults measured by a heated microcathode oxygen electrode.Dan Med Bull. 1988 Aug;35(4):322-34. Dan Med Bull. 1988. PMID: 3048920 Review.
-
Skin oxygen permeability in premature infants.Pediatrics. 1978 Oct;62(4):488-91. Pediatrics. 1978. PMID: 714579
-
Dermal heat transport analysis for transcutaneous O2 measurement.Acta Anaesthesiol Scand Suppl. 1978;68:4-8. Acta Anaesthesiol Scand Suppl. 1978. PMID: 279199 No abstract available.
-
Skin oxygen tension, skin oxygen consumption, and skin blood flow measured by a tc-pO2 electrode.Acta Physiol Scand Suppl. 1991;603:53-7. Acta Physiol Scand Suppl. 1991. PMID: 1789131 Review.
Cited by
-
New transcutaneous pO2 probe based on mass spectrometry with low stirring effect.Med Biol Eng Comput. 1985 May;23(3):254-7. doi: 10.1007/BF02446867. Med Biol Eng Comput. 1985. PMID: 4021570 No abstract available.
-
Novel transcutaneous sensor combining optical tcPO2 and electrochemical tcPCO2 monitoring with reflectance pulse oximetry.Med Biol Eng Comput. 2020 Feb;58(2):239-247. doi: 10.1007/s11517-019-02067-x. Epub 2019 Nov 18. Med Biol Eng Comput. 2020. PMID: 31741291 Free PMC article.
-
Transcutaneous PO2 measurement.Can Anaesth Soc J. 1984 Nov;31(6):664-77. doi: 10.1007/BF03008765. Can Anaesth Soc J. 1984. PMID: 6388743 Review.
-
Characterisation of steady-state and transient responses of transcutaneous pO2 and pCO2 by mass spectrometry.Med Biol Eng Comput. 1985 Nov;23(6):572-8. doi: 10.1007/BF02455312. Med Biol Eng Comput. 1985. PMID: 3935887 No abstract available.
MeSH terms
Substances
LinkOut - more resources
Research Materials
Miscellaneous