Shear wave elastography of the testes in a healthy study collective - Differences in standard values between ARFI and VTIQ techniques
- PMID: 27935546
- DOI: 10.3233/CH-168052
Shear wave elastography of the testes in a healthy study collective - Differences in standard values between ARFI and VTIQ techniques
Abstract
Introduction: Shear wave elastography (SWE) is a recent technique in the assessment of tissue elasticity. Different elastography techniques have been described over the years.Acoustic Radiation Force Impulse Imaging (ARFI) uses mechanical excitation of tissue to create detectable shear waves, a higher shear wave velocity being associated with an increased tissue stiffness.The Virtual Touch Tissue Imaging Quantification (VTIQ) method uses a mechanical push pulse as well, additionally creating a colour-coded map, in which tissue stiffness can be measured within a stored map on the ultrasound device after measurement.ARFI has been used in determining standard values in testes of a healthy study collective, VTIQ has already been used in the evaluation of unclear scrotal masses.Both techniques allow an operator-independent examination without application of mechanical pressure. The aim of this study was to determine whether there is a statistically significant difference between shear wave velocity values of both techniques in a healthy collective.
Material and methods: Twenty patients without known testicular pathology underwent standard B-mode sonography and additional shear wave elastography in the ARFI mode as well as the VTIQ technique of both testes using the Siemens Acuson S2000™ and S3000™ ultrasound devices (Siemens HealthCare, Erlangen, Germany).Measurements of shear wave velocity were performed in the upper pole, the central portion and the lower pole separately for each testis. Values were described in m/s.Statistical evaluation was performed using paired t-test analysis.
Results: We measured a mean shear wave velocity of 0.81 m/s using ARFI and 1.07 m/s with VTIQ.Shear wave velocities determined by VTIQ were all significantly higher than values gained in the ARFI mode. (p < 0.001 to p = 0.007). Values were between 0.22 and 0.29 m/s higher, when the examination was performed using VTIQ.
Conclusion: ARFI and VTIQ elastography modes both proved to be feasible techniques in the assessment of testicular tissue elasticity. Consideration of higher values for VTIQ is important, when different elastography measurement results are compared, especially for the application of devices in a clinical setting, e.g. in the work-up of scrotal masses.A calculable factor for a comparison between both devices is desirable, but to be further assessed in largerstudies.
Keywords: ARFI; Testis; VTIQ; shear wave elastography; ultrasonography.
Similar articles
-
The role of VTIQ as a new tissue strain analytics measurement technique in testicular lesions.Clin Hemorheol Microcirc. 2014;58(1):195-209. doi: 10.3233/CH-141904. Clin Hemorheol Microcirc. 2014. PMID: 25267458
-
Differences of standard values of Supersonic shear imaging and ARFI technique - in vivo study of testicular tissue.Clin Hemorheol Microcirc. 2016;64(4):729-733. doi: 10.3233/CH-168039. Clin Hemorheol Microcirc. 2016. PMID: 27792000
-
Three-dimensional vs. two-dimensional shear-wave elastography of the testes - preliminary study on a healthy collective.Clin Hemorheol Microcirc. 2016;64(3):447-456. doi: 10.3233/CH-168115. Clin Hemorheol Microcirc. 2016. PMID: 27886002
-
Association between diagnostic efficacy of acoustic radiation force impulse for benign and malignant thyroid nodules and the presence or absence of non-papillary thyroid cancer: A meta-analysis.Front Oncol. 2023 Jan 27;13:1007464. doi: 10.3389/fonc.2023.1007464. eCollection 2023. Front Oncol. 2023. PMID: 36776305 Free PMC article. Review.
-
Shear-Wave Elastography: Basic Physics and Musculoskeletal Applications.Radiographics. 2017 May-Jun;37(3):855-870. doi: 10.1148/rg.2017160116. Radiographics. 2017. PMID: 28493799 Free PMC article. Review.
Cited by
-
Shear Wave Elastography Evaluation of Testicular Stiffness in Dogs Affected by Testicular Pathology.Animals (Basel). 2025 Jan 26;15(3):353. doi: 10.3390/ani15030353. Animals (Basel). 2025. PMID: 39943123 Free PMC article.
-
[Sonoelastography in daily clinical routine].Radiologe. 2017 May;57(5):356-365. doi: 10.1007/s00117-017-0224-0. Radiologe. 2017. PMID: 28213851 Review. German.
-
Elastography in the Urological Practice: Urinary and Male Genital Tract, Prostate Excluded-Review.Diagnostics (Basel). 2022 Jul 16;12(7):1727. doi: 10.3390/diagnostics12071727. Diagnostics (Basel). 2022. PMID: 35885631 Free PMC article. Review.
-
Assessment of testicular stiffness in fertile dogs with shear wave elastography techniques: a pilot study.Front Vet Sci. 2024 May 2;11:1397347. doi: 10.3389/fvets.2024.1397347. eCollection 2024. Front Vet Sci. 2024. PMID: 38756525 Free PMC article.
-
Comparison of Tissue Stiffness Using Shear Wave Elastography in Men with Normal Testicular Tissue, Testicular Microlithiasis and Testicular Cancer.Ultrasound Int Open. 2017 Sep;3(4):E150-E155. doi: 10.1055/s-0043-116660. Epub 2017 Dec 20. Ultrasound Int Open. 2017. PMID: 29276802 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical