Role of vessel-to-prosthesis size mismatch in venous valve performance
- PMID: 27987598
- DOI: 10.1016/j.jvsv.2016.08.004
Role of vessel-to-prosthesis size mismatch in venous valve performance
Abstract
Background: Efforts to treat chronic venous insufficiency have focused on the development of prosthetic venous valves. The role of prosthetic valve-to-vessel size matching has not been determined. The purpose of this investigation was to assess the effect of size mismatching on venous valve function and to establish a mismatch limit that affects valve hemodynamic performance and venous wall stress to improve future valve designs and implants.
Methods: Flow dynamics of prosthetic venous valves were studied in vitro using a pulse duplicator flow loop. Valve performance based on flow rate and pressure measurements was determined at oversizing ratios ranging from 4.2% to 25%. Valve open area ratios at different size mismatching ratios were investigated by image analysis. Finally, a wall stress analysis was used to determine the magnitude of circumferential (hoop) stress in the venous wall at various degrees of oversizing.
Results: Our findings indicate that valve regurgitate volume, closing time, and pressure difference across the valve are significantly elevated at mismatch ratios greater than ∼15%. This is supported by increases in regurgitate velocity and open area relative to valves tested at near-nominal diameters. At this degree of size mismatch, the wall stress is increased by a factor of two to three times relative to physiologic pressures.
Conclusions: These findings establish a relationship between valve size matching and valve hemodynamic performance, including vessel wall stress, which should be considered in future valve implants. The size of the prosthetic valve should be within 15% of maximum vein size to optimize venous valve hemodynamic performance and to minimize the hoop wall stress.
Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Hemodynamics of venous valve pairing and implications on helical flow.J Vasc Surg Venous Lymphat Disord. 2018 Jul;6(4):517-522.e1. doi: 10.1016/j.jvsv.2018.02.005. J Vasc Surg Venous Lymphat Disord. 2018. PMID: 29909857
-
Characterization of a bioprosthetic bicuspid venous valve hemodynamics: implications for mechanism of valve dynamics.Eur J Vasc Endovasc Surg. 2014 Oct;48(4):459-64. doi: 10.1016/j.ejvs.2014.06.034. Epub 2014 Aug 21. Eur J Vasc Endovasc Surg. 2014. PMID: 25150441
-
Role of sinus in prosthetic venous valve.Eur J Vasc Endovasc Surg. 2014 Jul;48(1):98-104. doi: 10.1016/j.ejvs.2014.03.041. Epub 2014 May 10. Eur J Vasc Endovasc Surg. 2014. PMID: 24820915
-
Percutaneous venous valve designs for treatment of deep venous insufficiency.J Endovasc Ther. 2012 Apr;19(2):291-302. doi: 10.1583/11-3766R.1. J Endovasc Ther. 2012. PMID: 22545897 Review.
-
Historical overview of venous valve prostheses for the treatment of deep venous valve insufficiency.J Endovasc Ther. 2012 Apr;19(2):281-90. doi: 10.1583/11-3594MR.1. J Endovasc Ther. 2012. PMID: 22545896 Review.
Cited by
-
Advances in the diagnosis and treatment of primary deep venous valve insufficiency.Eur J Med Res. 2025 May 23;30(1):409. doi: 10.1186/s40001-025-02589-1. Eur J Med Res. 2025. PMID: 40410866 Free PMC article. Review.
-
Surgery for deep venous insufficiency.Cochrane Database Syst Rev. 2021 Sep 30;9(9):CD001097. doi: 10.1002/14651858.CD001097.pub4. Cochrane Database Syst Rev. 2021. PMID: 34591328 Free PMC article.
-
Transcatheter bicuspid venous valve prostheses: fluid mechanical performance testing of artificial nonwoven leaflets.Biomed Eng Online. 2024 Nov 29;23(1):124. doi: 10.1186/s12938-024-01316-x. Biomed Eng Online. 2024. PMID: 39614321 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources