A randomised comparison of three different immobilisation devices for thoracic and abdominal cancers

Affiliations
Comparative Study

A randomised comparison of three different immobilisation devices for thoracic and abdominal cancers

Catherine Hubie et al. J Med Radiat Sci. 2017 Jun.

Abstract

Introduction: Patient immobilisation is critically important for both highly conformal conventionally fractionated radiotherapy and for stereotactic body radiotherapy. Different immobilisation devices are available to maintain patient position for radiotherapy but the most suitable one remains unknown.

Methods: Forty-five patients were randomly allocated to one of three immobilisation devices; the Q fix arm shuttle, BodyFIX without wrap or BodyFIX with wrap. Patients were imaged before and after treatment to ascertain intra-fraction and inter-fraction motion. Bony anatomy was used for matching to determine the positional accuracy of each device. Treatments were timed using a standard method. Patient comfort and staff satisfaction questionnaires were also issued to determine comfort, ease of use and preferences for each device.

Results: The BodyFIX without wrap was the more accurate device; however, the differences between the devices were not statistically significant. The BodyFIX with wrap was found to take significantly longer to set up and set down compared to the arm shuttle and the BodyFIX without wrap (all P < 0.001). Patients (37%) marginally preferred the BodyFIX with wrap. Most (81%) staff preferred the BodyFIX without wrap.

Conclusion: Immobilisation using the BodyFIX without wrap was deemed to be suitable for clinical use. It was a clinically accurate device, the more efficient in terms of set up and set down time, the most preferred by staff and was accepted by patients.

Keywords: Abdomen; immobilisation; lung; radiotherapy.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Photographs of the three devices (from Left to right): Q‐Fix Arm shuttle, BodyFIX without vacuum wrap and BodyFIX with vacuum wrap.

References

    1. Dahele M, Verbakel W, Cuijpers J, Slotman B, Senan S. An analysis of patient positioning during stereotactic lung radiotherapy performed without rigid external immobilization. Radiother Oncol 2012; 104: 28–32. - PubMed
    1. Chang J, Balter P, Dong L, et al. Stereotactic body radiation therapy in centrally and superiorly located stage I or isolated recurrent non‐small‐cell lung cancer. Int J Radiat Oncol Biol Phys 2008; 72: 967–71. - PMC - PubMed
    1. Sahgal A, Roberge D, Schellenberg D, et al. The Canadian Association of radiation oncology scope of practice guidelines for lung, liver and spine stereotactic body radiotherapy. Clin Oncol 2012; 24: 629–39. - PubMed
    1. Winnie L, Purdie TG, Taremi M, et al. Effect of immobilization and performance status on intrafraction motion for stereotactic lung radiotherapy: Analysis of 133 patients. Int J Radiat Oncol Biol Phys 2011; 81: 1568–75. - PubMed
    1. Winnie L, Sahgal A, Foote M, Millar B‐A, Jaffray DA, Letoureau D. Impact of immobilization on intrafraction motion for spine stereotactic body radiotherapy using cone beam computed tomography. Int J Radiat Oncol Biol Phys 2012; 84: 520–6. - PubMed

Publication types

MeSH terms

LinkOut - more resources