Oxygen permeability of the pigmented material used in cosmetic daily disposable contact lenses
- PMID: 28003735
- PMCID: PMC5161396
- DOI: 10.2147/OPTH.S105222
Oxygen permeability of the pigmented material used in cosmetic daily disposable contact lenses
Abstract
Purpose: To evaluate the individual contributions of pigment colorant and packing solution containing polyvinyl pyrrolidone (PVP) on the oxygen permeability (Dk) of a cosmetic printed etafilcon A daily disposable contact lens packaged with PVP.
Method: The oxygen transport of a contact lens is evaluated through the central optical zone of the lens. Cosmetic printed contact lenses contain pigment colorant in the periphery or mid-periphery of the lens. Therefore, to assess the impact of cosmetic print on oxygen permeability, special lenses need to be produced that contain the colorant within the central optical zone. This technique was used to obtain multiple measurements of nonedge-corrected Dk/t of both the center pigmented lens and its nonpigmented equivalent, using a polarographic measurement described in International Organization for Standardization (ISO) 18369-4:2006(E), and the Dk derived for each measurement is corrected for edge effect. In addition, the edge-corrected Dk values of lenses made from the same monomer batch were measured. The lenses were packaged and autoclaved with and without proprietary technology which embeds PVP in the contact lens during autoclaving. The resulting Dk value of the printed lens material was then used with thickness data to generate true Dk/t profiles for a given lens power.
Results: The edge-corrected Dk of the printed etafilcon A lens with offset pigment colorant was measured to be 19.7×10-11 (cm2/s) (mL O2/mL·mmHg) at 35°C. This was within ±20% tolerance range as specified in ISO 18369-2:2012(E) for the edge-corrected Dk of the nonpigmented etafilcon A control lens evaluated during the same session, 19.5×10-11 (cm2/s) (mL O2/mL·mmHg). The edge-corrected Dk values of the lenses packaged with PVP (mean 20.1, standard deviation [SD] 0.3) were also within the ±20% tolerance range compared to those packaged without PVP (mean 20.0, SD 0.3).
Conclusion: The pigment colorant and PVP embedded in the contact lens during autoclaving were not found to influence the oxygen permeability of the etafilcon A material.
Keywords: Dk; cosmetic contact lenses; hydrogel contact lenses; oxygen; oxygen permeability.
Figures



Similar articles
-
Local oxygen transmissibility of disposable contact lenses.Cont Lens Anterior Eye. 2003 Dec;26(4):189-96. doi: 10.1016/j.clae.2003.09.001. Cont Lens Anterior Eye. 2003. PMID: 16303517
-
Oxygen permeability of disposable soft contact lenses.Am J Ophthalmol. 1990 Sep 15;110(3):269-73. doi: 10.1016/s0002-9394(14)76343-3. Am J Ophthalmol. 1990. PMID: 2396652
-
Oxygen permeability of soft contact lenses in different pH, osmolality and buffering solution.Int J Ophthalmol. 2015 Oct 18;8(5):1037-42. doi: 10.3980/j.issn.2222-3959.2015.05.33. eCollection 2015. Int J Ophthalmol. 2015. PMID: 26558223 Free PMC article.
-
Central Corneal Edema with Scleral-Lens Wear.Curr Eye Res. 2018 Nov;43(11):1305-1315. doi: 10.1080/02713683.2018.1500610. Epub 2018 Aug 6. Curr Eye Res. 2018. PMID: 30015536 Review.
-
Therapeutic contact lenses: the role of high-Dk lenses.Ophthalmol Clin North Am. 2003 Sep;16(3):455-61. doi: 10.1016/s0896-1549(03)00045-2. Ophthalmol Clin North Am. 2003. PMID: 14564766 Review.
Cited by
-
Influence of Selected Ophthalmic Fluids on the Wettability and Hydration of Hydrogel and Silicone Hydrogel Contact Lenses-In Vitro Study.Materials (Basel). 2022 Jan 25;15(3):930. doi: 10.3390/ma15030930. Materials (Basel). 2022. PMID: 35160876 Free PMC article.
-
ToF-SIMS and AFM Characterization of Brown Cosmetic Contact Lenses: From Structural Analysis to the Identification of Pigments.J Anal Methods Chem. 2020 Jan 22;2020:6134627. doi: 10.1155/2020/6134627. eCollection 2020. J Anal Methods Chem. 2020. PMID: 32399310 Free PMC article.
-
Experience of applying cosmetic Etafilcon A contact lens in cases with microcornea.Arq Bras Oftalmol. 2021 Jul 14;84(5):462-466. doi: 10.5935/0004-2749.20210075. eCollection 2021. Arq Bras Oftalmol. 2021. PMID: 34320106 Free PMC article.
-
Burkholderia Cepacia-Induced Bilateral Microbial Keratitis in a Wearer of Colored Cosmetic Contact Lenses.Cureus. 2022 Mar 4;14(3):e22840. doi: 10.7759/cureus.22840. eCollection 2022 Mar. Cureus. 2022. PMID: 35382181 Free PMC article.
References
-
- Erickson P, Comstock TL, Zantos SG. Effects of hydrogel lens transmissibility profiles on local corneal swelling during eye closure. Optom Vis Sci. 1996;73:169–177. - PubMed
-
- Papas EB. The role of hypoxia in the limbal vascular response to soft contact lens wear. Eye Cont Lens. 2003;29:S72–S74. discussion S83–S84, S192–S194. - PubMed
-
- Holden BA, Sweeney DF, Vannas A, Nilsson KT, Efron N. Effects of long-term extended contact lens wear on the human cornea. Invest Ophthalmol Vis Sci. 1985;26:1489–1501. - PubMed
-
- Holden BA, Sweeney DF, Swarbrick HA, et al. The vascular response to long-term extended contact lens wear. Clin Exp Optom. 1986;69:112–119. - PubMed
-
- Perez JG, Meijome JM, Jalbert I, Sweeney DF, Erickson P. Corneal epithelial thinning profile induced by long-term wear of hydrogel lenses. Cornea. 2003;22:304–307. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources