Comparison of the fit accuracy of zirconia-based prostheses generated by two CAD/CAM systems
- PMID: 28018561
- PMCID: PMC5179482
- DOI: 10.4047/jap.2016.8.6.439
Comparison of the fit accuracy of zirconia-based prostheses generated by two CAD/CAM systems
Abstract
Purpose: The purposes of this study are to evaluate the internal and marginal adaptation of two widely used CAD/CAM systems and to study the effect of porcelain press veneering process on the prosthesis adaptation.
Materials and methods: Molar of a lower jaw typodont resin model was prepared by adjusting a 1.0 mm circumferential chamfer, an occlusal reduction of 2.0 mm, and a 5° convergence angle and was duplicated as an abrasion-resistant master die. The monolithic crowns and copings were fabricated with two different CAD/CAM system-Ceramil and Zirkonzahn systems. Two kinds of non-destructive analysis methods are used in this study. First, weight technique was used to determine the overall fitting accuracy. And, to evaluate internal and marginal fit of specific part, replica technique procedures were performed.
Results: The silicone weight for the cement space of monolithic crowns and copings manufactured with Ceramil system was significantly higher than that from Zirkonzahn system. This gap might cause the differences in the silicone weight because the prostheses were manufactured according to the recommendation of each system. Marginal discrepancies of copings made with Ceramil system were between 106 and 117 µm and those from Zirkonzahn system were between 111 and 115 µm. Marginal discrepancies of copings made with Ceramil system were between 101 and 131 µm and those from Zirkonzahn system were between 116 and 131 µm.
Conclusion: Marginal discrepancy was relatively lower in Ceramil system and internal gap was smaller in Zirkonzahn system. There were significant differences in the internal gap of monolithic crown and coping among the 2 CAD/CAM systems. Marginal discrepancy produced from the 2 CAD/CAM systems were within a reported clinically acceptable range of marginal discrepancy.
Keywords: Fit accuracy; Press-over technique; Replica technique; Weight technique.
Figures





Similar articles
-
Comparison between five CAD/CAM systems for fit of zirconia copings.Quintessence Int. 2018;49(6):437-444. doi: 10.3290/j.qi.a40354. Quintessence Int. 2018. PMID: 29756132
-
Effect of feldspathic porcelain layering on the marginal fit of zirconia and titanium complete-arch fixed implant-supported frameworks.J Prosthet Dent. 2018 Jul;120(1):71-78. doi: 10.1016/j.prosdent.2017.11.003. Epub 2018 Feb 6. J Prosthet Dent. 2018. PMID: 29426786
-
Influence of porcelain veneering on the marginal fit of Digident and Lava CAD/CAM zirconia ceramic crowns.J Adv Prosthodont. 2010 Jun;2(2):33-8. doi: 10.4047/jap.2010.2.2.33. Epub 2010 Jun 30. J Adv Prosthodont. 2010. PMID: 21165185 Free PMC article.
-
The effect of different fabrication steps on the marginal adaptation of two types of glass-infiltrated ceramic crown copings fabricated by CAD/CAM technology.J Prosthodont. 2012 Apr;21(3):167-72. doi: 10.1111/j.1532-849X.2011.00803.x. Epub 2012 Feb 28. J Prosthodont. 2012. PMID: 22372838
-
Assessment Methods for Marginal and Internal Fit of Partial Crown Restorations: A Systematic Review.J Clin Med. 2023 Jul 31;12(15):5048. doi: 10.3390/jcm12155048. J Clin Med. 2023. PMID: 37568450 Free PMC article. Review.
Cited by
-
Effect of virtual cement space and restorative materials on the adaptation of CAD-CAM endocrowns.BMC Oral Health. 2022 Dec 9;22(1):580. doi: 10.1186/s12903-022-02598-0. BMC Oral Health. 2022. PMID: 36494663 Free PMC article.
-
Influence of Material Selection on the Marginal Accuracy of CAD/CAM-Fabricated Metal- and All-Ceramic Single Crown Copings.Biomed Res Int. 2018 Mar 22;2018:2143906. doi: 10.1155/2018/2143906. eCollection 2018. Biomed Res Int. 2018. PMID: 29765979 Free PMC article.
-
Fracture resistance and marginal fit of the zirconia crowns with varied occlusal thickness.J Adv Prosthodont. 2020 Oct;12(5):283-290. doi: 10.4047/jap.2020.12.5.283. Epub 2020 Oct 26. J Adv Prosthodont. 2020. PMID: 33149849 Free PMC article.
-
Comparative evaluation of internal fit and marginal gap of endocrowns using lithium disilicate and polyether ether ketone materials - an in vitro study.BMC Oral Health. 2023 Apr 7;23(1):207. doi: 10.1186/s12903-023-02857-8. BMC Oral Health. 2023. PMID: 37029396 Free PMC article.
-
An Advanced Fiber-Reinforced Composite Solution for Gingival Inflammation and Bone Loss Related to Restorative Crowns.EC Dent Sci. 2020 Feb;19(2):https://www.ecronicon.com/ecde/pdf/ECDE-19-01347.pdf. Epub 2020 Jan 29. EC Dent Sci. 2020. PMID: 33196061 Free PMC article.
References
-
- Hannink RHJ, Kelly PM, Muddle BC. Transformation toughening in zirconia-containing ceramics. J Am Ceram Soc. 2000;83:461–487.
-
- Conrad HJ, Seong WJ, Pesun IJ. Current ceramic materials and systems with clinical recommendations: a systematic review. J Prosthet Dent. 2007;98:389–404. - PubMed
-
- Baig MR, Tan KB, Nicholls JI. Evaluation of the marginal fit of a zirconia ceramic computer-aided machined (CAM) crown system. J Prosthet Dent. 2010;104:216–227. - PubMed
-