Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2016 Nov 28;9(22):2292-2303.
doi: 10.1016/j.jcin.2016.08.037. Epub 2016 Oct 26.

Comparative Hemodynamic Effects of Contemporary Percutaneous Mechanical Circulatory Support Devices in a Porcine Model of Acute Myocardial Infarction

Affiliations
Comparative Study

Comparative Hemodynamic Effects of Contemporary Percutaneous Mechanical Circulatory Support Devices in a Porcine Model of Acute Myocardial Infarction

Brian R Weil et al. JACC Cardiovasc Interv. .

Abstract

Objectives: The aim of this study was to directly compare the hemodynamic effects of 2 contemporary percutaneous mechanical circulatory support devices in a porcine model of acute myocardial infarction.

Background: Percutaneous support devices offer the ability to unload the ischemic left ventricle, but the comparative hemodynamic effects of contemporary platforms are unclear.

Methods: Yorkshire swine (mean weight 76 ± 2 kg; n = 7) were instrumented with a left ventricular (LV) pressure-volume (PV) catheter and subjected to a 2-h coronary occlusion. Hemodynamic parameters and PV-derived indexes of LV performance were assessed 30 min after reperfusion and during LV support with Impella CP (ICP) and TandemHeart devices (in randomized order) at comparable flow rates.

Results: Myocardial infarction produced a rightward shift of the PV loop and increased LV end-diastolic pressure (from 9 ± 2 mm Hg to 15 ± 2 mm Hg; p = 0.04). After reperfusion, both devices maintained aortic pressure, shifted the PV loop to the left, and decreased LV end-diastolic pressure (ICP vs. TandemHeart; 11 ± 1 mm Hg vs. 7 ± 4 mm Hg; p = 0.04). However, only TandemHeart elicited significant reductions in native LV stroke volume (from 75 ± 7 ml to 39 ± 7 ml; p < 0.01), dP/dtmax (from 988 ± 77 mm Hg/s to 626 ± 42 mm Hg/s; p < 0.01), stroke work (from 0.70 ± 0.03 J to 0.26 ± 0.05 J; p < 0.01), PV area (from 0.95 ± 0.11 J to 0.47 ± 0.10 J; p < 0.01), and pre-load-recruitable stroke work slope (from 41.7 ± 2.8 J/ml to 30.6 ± 3.9 J/ml; p = 0.05).

Conclusions: At comparable device flow rates, TandemHeart decreased LV pre-load, native LV stroke volume, and myocardial contractility to a greater degree than ICP. Reductions in load-independent indexes of LV performance indicate favorable effects on myocardial oxygen balance and support further study of TandemHeart in clinical scenarios requiring mechanical support in the setting of acute myocardial ischemia.

Keywords: acute myocardial infarction; left ventricular assist device; percutaneous mechanical circulatory support; ventricular unloading.

PubMed Disclaimer

Figures

FIGURE 1
FIGURE 1. Experimental Protocol
Following baseline data collection, the left circumflex coronary artery was occluded for 2 h. Thirty min after reperfusion, data collection was repeated and animals were randomized to mechanical circulatory support with either Impella CP (ICP) or TandemHeart. After data collection, the first device was removed and data collection was repeated during a second unassisted post–myocardial infarction (MI) time point, followed by insertion of the second circulatory support device. PV = pressure-volume; TTC = triphenyl tetrazolium chloride.
FIGURE 2
FIGURE 2. Porcine Model of Acute Myocardial Infarction
(A) Example fluoroscopic images illustrating the percutaneous coronary occlusion procedure. (B) Data collected 30 min after reperfusion demonstrated a rightward shift in the pressure-volume (PV) loop, an elevation in left ventricular (LV) end-diastolic pressure, and a reduction in LV ejection fraction. (C) PV loops collected during rapid inferior vena cava occlusion permitted assessment of load-independent parameters of LV performance. Compared with baseline, end-systolic elastance (the slope of the end-systolic PV relationship) was decreased following myocardial infarction (MI). (D) Following excision, the left ventricle was sectioned into 5 short-axis slices for post-mortem assessment of infarct size. LCx = left circumflex coronary artery.
FIGURE 2
FIGURE 2. Porcine Model of Acute Myocardial Infarction
(A) Example fluoroscopic images illustrating the percutaneous coronary occlusion procedure. (B) Data collected 30 min after reperfusion demonstrated a rightward shift in the pressure-volume (PV) loop, an elevation in left ventricular (LV) end-diastolic pressure, and a reduction in LV ejection fraction. (C) PV loops collected during rapid inferior vena cava occlusion permitted assessment of load-independent parameters of LV performance. Compared with baseline, end-systolic elastance (the slope of the end-systolic PV relationship) was decreased following myocardial infarction (MI). (D) Following excision, the left ventricle was sectioned into 5 short-axis slices for post-mortem assessment of infarct size. LCx = left circumflex coronary artery.
FIGURE 3
FIGURE 3. Hemodynamic Effects of Impella CP and TandemHeart Following Acute Myocardial Infarction
(A) Example fluoroscopic images demonstrating anatomic placement of each device. (B) Representative pressure-volume (PV) loops illustrate the hemodynamic effects of each device (blue Impella CP and green TandemHeart) compared with the corresponding unassisted post–myocardial infarction (MI) baseline time point (red). Both devices shifted the PV loop to the left and decreased left ventricular (LV) end-diastolic pressure, but only TandemHeart (TH) produced a reduction in native LV stroke volume. (C) The leftward shift of the PV loop with each device is quantitatively illustrated by the reduction in LV end-diastolic volume. However, LV end-systolic volume differed during support with the 2 devices, such that only TH produced a significant reduction in native LV stroke volume. As a result, native LV cardiac output decreased only during TH support despite similar pulmonary blood flow (i.e., total cardiac output) with both devices. (D) TH produced greater reductions in LV dP/dtmax and stroke work compared with Impella CP (ICP), indicative of decreased myocardial contractility and mechanical work during TH unloading. *p < 0.05 versus post-MI; †p < 0.05 versus ICP.
FIGURE 3
FIGURE 3. Hemodynamic Effects of Impella CP and TandemHeart Following Acute Myocardial Infarction
(A) Example fluoroscopic images demonstrating anatomic placement of each device. (B) Representative pressure-volume (PV) loops illustrate the hemodynamic effects of each device (blue Impella CP and green TandemHeart) compared with the corresponding unassisted post–myocardial infarction (MI) baseline time point (red). Both devices shifted the PV loop to the left and decreased left ventricular (LV) end-diastolic pressure, but only TandemHeart (TH) produced a reduction in native LV stroke volume. (C) The leftward shift of the PV loop with each device is quantitatively illustrated by the reduction in LV end-diastolic volume. However, LV end-systolic volume differed during support with the 2 devices, such that only TH produced a significant reduction in native LV stroke volume. As a result, native LV cardiac output decreased only during TH support despite similar pulmonary blood flow (i.e., total cardiac output) with both devices. (D) TH produced greater reductions in LV dP/dtmax and stroke work compared with Impella CP (ICP), indicative of decreased myocardial contractility and mechanical work during TH unloading. *p < 0.05 versus post-MI; †p < 0.05 versus ICP.
FIGURE 4
FIGURE 4. Load-Independent Indexes of Left Ventricular Performance During Percutaneous Mechanical Circulatory Support
(A) Representative pressure-volume (PV) loops collected with transient pre-load reduction following reperfused myocardial infarction (MI) (left) with illustration of PV area (PVA) calculation (right). PVA is the product of stroke work (SW) (mechanical energy) and residual (potential) energy and provides an index of myocardial oxygen consumption. (B) Representative PV loops collected with transient pre-load reduction during mechanical circulatory support with Impella CP (ICP) (left) and TandemHeart (TH) (right). (C) Load-independent indexes of myocardial oxygen consumption (PVA) and contractility (pre-load-recruitable SW) were each significantly reduced during TH support. *p < 0.05 versus post-MI; †p < 0.05 versus ICP.
FIGURE 4
FIGURE 4. Load-Independent Indexes of Left Ventricular Performance During Percutaneous Mechanical Circulatory Support
(A) Representative pressure-volume (PV) loops collected with transient pre-load reduction following reperfused myocardial infarction (MI) (left) with illustration of PV area (PVA) calculation (right). PVA is the product of stroke work (SW) (mechanical energy) and residual (potential) energy and provides an index of myocardial oxygen consumption. (B) Representative PV loops collected with transient pre-load reduction during mechanical circulatory support with Impella CP (ICP) (left) and TandemHeart (TH) (right). (C) Load-independent indexes of myocardial oxygen consumption (PVA) and contractility (pre-load-recruitable SW) were each significantly reduced during TH support. *p < 0.05 versus post-MI; †p < 0.05 versus ICP.

Comment in

References

    1. Mozaffarian D, Benjamin EJ, Go AS, et al. Heart disease and stroke statistics—2015 update: a report from the American Heart Association. Circulation. 2015;131:e29–322. - PubMed
    1. Jeger RV, Radovanovic D, Hunziker PR, et al. Ten-year trends in the incidence and treatment of cardiogenic shock. Ann Intern Med. 2008;149:618–26. - PubMed
    1. Goldberg RJ, Spencer FA, Gore JM, Lessard D, Yarzebski J. Thirty-year trends (1975–2005) in the magnitude of, management of, and hospital death rates associated with cardiogenic shock in patients with acute myocardial infarction: a population-based perspective. Circulation. 2009;119:1211–9. - PMC - PubMed
    1. Aissaoui N, Puymirat E, Tabone X, et al. Improved outcome of cardiogenic shock at the acute stage of myocardial infarction: a report from the USIK 1995, USIC 2000, and FAST-MI French nationwide registries. Eur Heart J. 2012;33:2535–43. - PubMed
    1. Thiele H, Zeymer U, Neumann FJ, et al. Intraaortic balloon support for myocardial infarction with cardiogenic shock. N Engl J Med. 2012;367:1287–96. - PubMed

Publication types