Evaluation of the Anterior Segment Angle-to-Angle Scan of Cirrus High-Definition Optical Coherence Tomography and Comparison With Gonioscopy and With the Visante OCT
- PMID: 28061511
- DOI: 10.1167/iovs.16-20886
Evaluation of the Anterior Segment Angle-to-Angle Scan of Cirrus High-Definition Optical Coherence Tomography and Comparison With Gonioscopy and With the Visante OCT
Abstract
Purpose: To evaluate the diagnostic performance of the anterior segment angle-to-angle scan of the Cirrus high-definition optical coherence tomography (HD-OCT) in detecting eyes with closed angles.
Methods: All subjects underwent dark-room gonioscopy by an ophthalmologist. A technician performed anterior segment imaging with Cirrus (n = 202) and Visante OCT (n = 85) under dark-room conditions. All eyes were categorized by two masked graders as per number of closed quadrants. Each quadrant of anterior chamber angle was categorized as a closed angle if posterior trabecular meshwork could not be seen on gonioscopy or if there was any irido-corneal contact anterior to scleral spur in Cirrus and Visante images. An eye was graded as having a closed angle if two or more quadrants were closed. Agreement and area under the curve (AUC) were performed.
Results: There were 50 (24.8%) eyes with closed angles. The agreements of closed-angle diagnosis (by eye) between Cirrus HD-OCT and gonioscopy (k = 0.59; 95% confidence interval (CI) 0.45-0.72; AC1 = 0.76) and between Cirrus and Visante OCT (k = 0.65; 95% CI 0.48-0.82, AC1 = 0.77) were moderate. The AUC for diagnosing the eye with gonioscopic closed angle by Cirrus HD-OCT was good (AUC = 0.86; sensitivity = 83.33; specificity = 77.78). The diagnostic performance of Cirrus HD-OCT in detecting the eyes with closed angles was similar to that of Visante (AUC 0.87 vs. 0.9, respectively; P = 0.51).
Conclusions: The anterior segment angle-to-angle scans of Cirrus HD-OCT demonstrated similar diagnostic performance as Visante in detecting gonioscopic closed angles. The agreement between Cirrus and gonioscopy for detecting eyes with closed angles was moderate.
Similar articles
-
Comparison of gonioscopy and anterior segment ocular coherence tomography in detecting angle closure in different quadrants of the anterior chamber angle.Ophthalmology. 2008 May;115(5):769-74. doi: 10.1016/j.ophtha.2007.06.030. Epub 2007 Oct 4. Ophthalmology. 2008. PMID: 17916377
-
Detection of primary angle closure using anterior segment optical coherence tomography in Asian eyes.Ophthalmology. 2007 Jan;114(1):33-9. doi: 10.1016/j.ophtha.2006.05.073. Epub 2006 Oct 27. Ophthalmology. 2007. PMID: 17070597
-
Comparing Gonioscopy With Visante and Cirrus Optical Coherence Tomography for Anterior Chamber Angle Assessment in Glaucoma Patients.J Glaucoma. 2016 Feb;25(2):177-83. doi: 10.1097/IJG.0000000000000076. J Glaucoma. 2016. PMID: 24844543
-
Role of anterior segment optical coherence tomography in angle-closure disease: a review.Clin Exp Ophthalmol. 2018 Mar;46(2):147-157. doi: 10.1111/ceo.13120. Epub 2017 Dec 28. Clin Exp Ophthalmol. 2018. PMID: 29193702 Review.
-
Development in anterior segment imaging for glaucoma.Curr Opin Ophthalmol. 2014 Mar;25(2):98-103. doi: 10.1097/ICU.0000000000000026. Curr Opin Ophthalmol. 2014. PMID: 24370972 Review.
Cited by
-
Diagnostic accuracy of AS-OCT vs gonioscopy for detecting angle closure: a systematic review and meta-analysis.Graefes Arch Clin Exp Ophthalmol. 2022 Jan;260(1):1-23. doi: 10.1007/s00417-021-05271-4. Epub 2021 Jul 5. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 34223989 Free PMC article.
-
Technologies for the diagnosis of angle closure glaucoma (ACE): protocol of a prospective, multicentre, cross-sectional diagnostic study.BMJ Open. 2023 Oct 4;13(10):e073975. doi: 10.1136/bmjopen-2023-073975. BMJ Open. 2023. PMID: 37793921 Free PMC article.
-
Practice Guidelines for Ocular Telehealth-Diabetic Retinopathy, Third Edition.Telemed J E Health. 2020 Apr;26(4):495-543. doi: 10.1089/tmj.2020.0006. Epub 2020 Mar 25. Telemed J E Health. 2020. PMID: 32209018 Free PMC article.
-
Telemedicine for Glaucoma: Guidelines and Recommendations.Telemed J E Health. 2020 Apr;26(4):551-555. doi: 10.1089/tmj.2020.0009. Epub 2020 Mar 25. Telemed J E Health. 2020. PMID: 32209001 Free PMC article.
-
Evaluation of Schlemm's canal with swept-source optical coherence tomography in primary angle-closure disease.BMC Ophthalmol. 2023 Jun 7;23(1):256. doi: 10.1186/s12886-023-03001-4. BMC Ophthalmol. 2023. PMID: 37286943 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources