One-step transvaginal three-dimensional hysterosalpingo-foam sonography (3D-HyFoSy) confirmation test for Essure® follow-up: a multicenter study
- PMID: 28067009
- DOI: 10.1002/uog.17398
One-step transvaginal three-dimensional hysterosalpingo-foam sonography (3D-HyFoSy) confirmation test for Essure® follow-up: a multicenter study
Abstract
Objective: To evaluate, in patients who underwent Fallopian-tube sterilization by hysteroscopic insertion of an Essure® device, the feasibility and accuracy of three-dimensional (3D) transvaginal sonography (TVS) to check the position of the device and 3D hysterosalpingo-foam sonography (3D-HyFoSy) using contrast-enhanced gel foam to assess consequent tubal occlusion.
Methods: This was a prospective multicenter study conducted from June 2012 to July 2014 in four Italian centers, which included 50 women who underwent hysteroscopic Essure microinsert placement in a total of 95 Fallopian tubes. Placement of the microinserts was performed in an outpatient setting following standard procedure recommendations. All patients underwent transvaginal 3D-HyFoSy and hysterosalpingography (HSG) approximately 12-14 weeks after the procedure. The position of the devices was first checked on 3D-TVS and classified according to specific criteria (Positions A, B, C and D). Then, 3D-HyFoSy with ExEm® gel foam as the ultrasound contrast agent was performed to confirm tubal occlusion by the absence of contrast agent within the tubes and/or around the ovaries. To evaluate the feasibility of 3D-HyFoSy, consecutive volume acquisitions were performed while injecting the gel foam. After sonographic evaluation, women underwent HSG to assess the success of sterilization, as standard methodology. Side effects and pain evoked during and after 3D-HyFoSy and HSG were evaluated using a numeric pain rating scale.
Results: On 3D-TVS, 10 devices (10.5%) were in Position A, two (2.1%) in Position B, 59 (62.1%) in Position C and 24 (25.3%) in Position D. During 3D-HyFoSy, tubal occlusion was observed in 89 of 95 tubes (93.7%). In the six cases in which the passage of the contrast was observed, one device (16.7%) was in Position B, one device (16.7%) in Position D and four devices (66.7%) were found to lie in Position C. Tubal patency was confirmed at HSG with a concordance rate of 100%. The mean pain score associated with 3D-HyFoSy was significantly lower than that on HSG.
Conclusions: 3D-TVS with HyFoSy allows accurate assessment of the position of Essure microinserts and tubal occlusion. It can be considered a safe, reliable, non-invasive alternative to HSG. Copyright © 2017 ISUOG. Published by John Wiley & Sons Ltd.
Keywords: Essure® device; HyFoSy; follow-up; hysteroscopic tubal sterilization; three-dimensional ultrasound.
Copyright © 2017 ISUOG. Published by John Wiley & Sons Ltd.
Similar articles
-
Use of Hysterosalpingo-Foam Sonography for Assessment of the Efficacy of Essure Hysteroscopic Sterilization.J Ultrasound Med. 2018 Aug;37(8):1929-1935. doi: 10.1002/jum.14539. Epub 2018 Jan 18. J Ultrasound Med. 2018. PMID: 29344973
-
Three-Dimensional Hysterosalpingo Contrast Sonography with Gel Foam: Methodology and Feasibility to Obtain 3-Dimensional Volumes of Tubal Shape.J Minim Invasive Gynecol. 2017 Jul-Aug;24(5):827-832. doi: 10.1016/j.jmig.2017.04.007. Epub 2017 Apr 24. J Minim Invasive Gynecol. 2017. PMID: 28450252
-
Diagnostic accuracy of hysterosalpingo-foam-sonography to confirm tubal occlusion after Essure® placement as treatment for hydrosalpinges.Reprod Biomed Online. 2015 Apr;30(4):421-5. doi: 10.1016/j.rbmo.2014.12.009. Epub 2015 Jan 7. Reprod Biomed Online. 2015. PMID: 25682307
-
Is hysterosalpingo-foam sonography the new gold standard for assessing tubal patency? A systematic review and meta-analysis.Reprod Biomed Online. 2025 Jan;50(1):104380. doi: 10.1016/j.rbmo.2024.104380. Epub 2024 Jul 18. Reprod Biomed Online. 2025. PMID: 39522344
-
Clinical Aspects of HyFoSy as Tubal Patency Test in Subfertility Workup.Biomed Res Int. 2019 Jul 8;2019:4827376. doi: 10.1155/2019/4827376. eCollection 2019. Biomed Res Int. 2019. PMID: 31360713 Free PMC article. Review.
Cited by
-
Follicular Fluid Metabolomics: Tool for Predicting IVF Outcomes of Different Infertility Causes.Reprod Sci. 2025 Apr;32(4):921-934. doi: 10.1007/s43032-024-01664-y. Epub 2024 Aug 1. Reprod Sci. 2025. PMID: 39090336 Free PMC article. Review.
Publication types