A comparison of two methods of quality control of pH, Pco2, and Po2 measurements
- PMID: 28069
- DOI: 10.1177/000456327801500137
A comparison of two methods of quality control of pH, Pco2, and Po2 measurements
Abstract
pH, Pco2, and Po2 measurements have been made over a three-week period on General Diagnostics Blood GAS Control Ampoules and on a tonometered buffer system. The latter seems adequate as a precision and accuracy control of pH, Pco2, and Po2 measurements in a single solution. The former seems adequate as a precision control of these measurements, but, on the particular batch provided, mean values need to be determined within the customer's laboratory.
Similar articles
-
Evaluation of ampouled tonometered buffer solutions as a quality-control system for pH, pCO2, and pO2 measurement.Clin Chem. 1977 Sep;23(9):1718-25. Clin Chem. 1977. PMID: 19168
-
Use of carbon dioxide- and oxygen-tonometered phosphate-bicarbonate-chloride-glycerol-water mixtures for calibration and control of pH, pCO2, and pO2 electrode systems.Clin Chem. 1975 May;21(6):685-93. Clin Chem. 1975. PMID: 235377
-
[Analytical reliability of the results of determining pH, pCO2 and pO2 with the ABL 300 (Radiometer) automatic pH and blood gas analyzer].Vutr Boles. 1988;27(4):77-81. Vutr Boles. 1988. PMID: 3145645 Bulgarian.
-
Capillary blood gases in the neonate. A reassessment and review of the literature.Am J Dis Child. 1990 Feb;144(2):168-72. doi: 10.1001/archpedi.1990.02150260046025. Am J Dis Child. 1990. PMID: 2105629 Review.
-
Temperature correction of blood-gas and pH measurements.Clin Chem. 1983 Nov;29(11):1877-85. Clin Chem. 1983. PMID: 6354511 Review.