Demonstration of blood flow by color doppler in the femoral artery distal to arterial cannula during peripheral venoarterial-extracorporeal membrane oxygenation
- PMID: 28074808
- PMCID: PMC5290680
- DOI: 10.4103/0971-9784.197848
Demonstration of blood flow by color doppler in the femoral artery distal to arterial cannula during peripheral venoarterial-extracorporeal membrane oxygenation
Abstract
In spite of distal perfusion of the limb using a cannula, the limb can have ischemic events if there is an undetected kink or clot anywhere in the line or thrombus in the artery. There are several ways to monitor and assess the limb ischemia. Monitoring for clinical signs of limb ischemia like temperature change and pallor is reliable and mandatory. We report a method where we used color Doppler to document the blood flow. Curvilinear vascular probe of an echo machine is used to identify the flow in the distal femoral artery of the lower limb. . As we have demonstrated in the video attached, once flow to the distal limb perfusion system is shut off by closing the three way stop cock, we can appreciate the immediate cessation of flow in the artery by Doppler.
Conflict of interest statement
There are no conflicts of interest.
References
-
- Bisdas T, Beutel G, Warnecke G, Hoeper MM, Kuehn C, Haverich A, et al. Vascular complications in patients undergoing femoral cannulation for extracorporeal membrane oxygenation support. Ann Thorac Surg. 2011;92:626–31. - PubMed
-
- Aziz F, Brehm CE, El-Banyosy A, Han DC, Atnip RG, Reed AB. Arterial complications in patients undergoing extracorporeal membrane oxygenation via femoral cannulation. Ann Vasc Surg. 2014;28:178–83. - PubMed
-
- Roussel A, Al-Attar N, Khaliel F, Alkhoder S, Raffoul R, Alfayyadh F, et al. Arterial vascular complications in peripheral extracorporeal membrane oxygenation support: A review of techniques and outcomes. Future Cardiol. 2013;9:489–95. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials