Analysing a mechanism of failure in retrieved magnetically controlled spinal rods

Affiliations

Analysing a mechanism of failure in retrieved magnetically controlled spinal rods

Vasiliki C Panagiotopoulou et al. Eur Spine J. 2017 Jun.

Abstract

Purpose: We aim to describe a mechanism of failure in magnetically controlled growth rods which are used for the correction of the early onset scoliosis.

Methods: This retrieval study involved nine magnetically controlled growth rods, of a single design, revised from five patients for metal staining, progression of scoliosis, swelling, fractured actuator pin, and final fusion. All the retrieved rods were radiographed and assessed macroscopically and microscopically for material loss. Two implants were further analysed using micro-CT scanning and then sectioned to allow examination of the internal mechanism. No funding was obtained to analyse these implants. There were no potential conflicts interests.

Results: Plain radiographs revealed that three out of nine retrieved rods had a fractured pin. All had evidence of surface degradation on the extendable telescopic rod. There was considerable corrosion along the internal mechanism.

Conclusions: We found that a third of the retrieved magnetically controlled growth rods had failed due to pin fracture secondary to corrosion of the internal mechanism. We recommend that surgeons consider that any inability of magnetically controlled growth rods to distract may be due to corrosive debris building up inside the mechanism, thereby preventing normal function.

Keywords: Corrosion; Early onset scoliosis; Magnetically controlled growth rods; Spine implants; Spine retrievals.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Spine (Phila Pa 1976). 2015 Jul 1;40(13):E767-73 - PubMed
    1. Spine J. 2016 Apr;16(4 Suppl):S34-9 - PubMed
    1. J Bone Joint Surg Am. 2010 Nov 3;92 (15):2533-43 - PubMed
    1. Appl Health Econ Health Policy. 2014 Dec;12 (6):587-99 - PubMed
    1. Spine (Phila Pa 1976). 2003 Oct 15;28(20):2397-406 - PubMed

LinkOut - more resources