Analysing a mechanism of failure in retrieved magnetically controlled spinal rods
- PMID: 28102447
- DOI: 10.1007/s00586-016-4936-z
Analysing a mechanism of failure in retrieved magnetically controlled spinal rods
Abstract
Purpose: We aim to describe a mechanism of failure in magnetically controlled growth rods which are used for the correction of the early onset scoliosis.
Methods: This retrieval study involved nine magnetically controlled growth rods, of a single design, revised from five patients for metal staining, progression of scoliosis, swelling, fractured actuator pin, and final fusion. All the retrieved rods were radiographed and assessed macroscopically and microscopically for material loss. Two implants were further analysed using micro-CT scanning and then sectioned to allow examination of the internal mechanism. No funding was obtained to analyse these implants. There were no potential conflicts interests.
Results: Plain radiographs revealed that three out of nine retrieved rods had a fractured pin. All had evidence of surface degradation on the extendable telescopic rod. There was considerable corrosion along the internal mechanism.
Conclusions: We found that a third of the retrieved magnetically controlled growth rods had failed due to pin fracture secondary to corrosion of the internal mechanism. We recommend that surgeons consider that any inability of magnetically controlled growth rods to distract may be due to corrosive debris building up inside the mechanism, thereby preventing normal function.
Keywords: Corrosion; Early onset scoliosis; Magnetically controlled growth rods; Spine implants; Spine retrievals.
Similar articles
-
Analysis of Explanted Magnetically Controlled Growing Rods From Seven UK Spinal Centers.Spine (Phila Pa 1976). 2018 Jan 1;43(1):E16-E22. doi: 10.1097/BRS.0000000000002221. Spine (Phila Pa 1976). 2018. PMID: 28459777
-
Retrieval and clinical analysis of distraction-based dual growing rod constructs for early-onset scoliosis.Spine J. 2017 Oct;17(10):1506-1518. doi: 10.1016/j.spinee.2017.04.020. Epub 2017 Apr 26. Spine J. 2017. PMID: 28456673
-
Metallosis following implantation of magnetically controlled growing rods in the treatment of scoliosis: a case series.Bone Joint J. 2016 Dec;98-B(12):1662-1667. doi: 10.1302/0301-620X.98B12.38061. Bone Joint J. 2016. PMID: 27909129
-
Systematic review of the complications associated with magnetically controlled growing rods for the treatment of early onset scoliosis.Eur Spine J. 2018 Sep;27(9):2062-2071. doi: 10.1007/s00586-018-5590-4. Epub 2018 Apr 19. Eur Spine J. 2018. PMID: 29675673
-
Magnetically controlled growing rods for scoliosis surgery.Expert Rev Med Devices. 2017 Feb;14(2):117-126. doi: 10.1080/17434440.2016.1274230. Epub 2017 Jan 3. Expert Rev Med Devices. 2017. PMID: 28044467 Review.
Cited by
-
Magnetically Controlled Growing Rods: The Experience of Mechanical Failure from a Single Center Consecutive Series of 28 Children with a Minimum Follow-up of 2 Years.Asian Spine J. 2018 Oct;12(5):794-802. doi: 10.31616/asj.2018.12.5.794. Epub 2018 Sep 10. Asian Spine J. 2018. PMID: 30213160 Free PMC article.
-
Identifying complications and failure modes of innovative growing rod configurations using the (hybrid) magnetically controlled growing rod (MCGR) and the spring distraction system (SDS).Spine Deform. 2021 Nov;9(6):1679-1689. doi: 10.1007/s43390-021-00378-4. Epub 2021 Jun 22. Spine Deform. 2021. PMID: 34156666 Free PMC article.
-
A retrieval analysis of the Precice intramedullary limb lengthening system.Bone Joint Res. 2018 Aug 4;7(7):476-484. doi: 10.1302/2046-3758.77.BJR-2017-0359.R1. eCollection 2018 Jul. Bone Joint Res. 2018. PMID: 30123497 Free PMC article.
-
Magnetically controlled growing rods in the treatment of early onset scoliosis: a single centre experience of 44 patients with mean follow-up of 4.1 years.Bone Jt Open. 2020 Nov 2;1(7):405-414. doi: 10.1302/2633-1462.17.BJO-2020-0099.R1. eCollection 2020 Jul. Bone Jt Open. 2020. PMID: 33215131 Free PMC article.
-
Mechanical wear analysis helps understand a mechanism of failure in retrieved magnetically controlled growing rods: a retrieval study.BMC Musculoskelet Disord. 2020 Aug 5;21(1):519. doi: 10.1186/s12891-020-03543-4. BMC Musculoskelet Disord. 2020. PMID: 32758204 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical