X-ray scatter correction for multi-source interior computed tomography
- PMID: 28102959
- DOI: 10.1002/mp.12022
X-ray scatter correction for multi-source interior computed tomography
Abstract
Purpose: The schemes of multi-source interior computed tomography (CT) have shown promise for ultra-fast, organ-oriented, and low-dose dynamic imaging. Besides forward scattering, x-ray cross scattering from multiple x-ray sources activated simultaneously can further degrade image quality. Here, we investigate the overall x-ray scattering artifact in a recently proposed multi-source interior CT architecture, and present two methods for scatter correction.
Methods: Compared to single-source global CT, scattering in multi-source interior CT architecture is affected by two new factors: cross scattering from simultaneously activated multiple x-ray sources and region-of-interest (ROI) oriented interior CT mode. The scatter artifact in the multi-source interior CT architecture was evaluated through both numerical simulation and physical experimentation, and compared to that from conventional single-source global CT. Monte Carlo simulation was conducted with a modified numerical CATphan® 600 phantom. Physical experiments were performed in an in-house developed CT imaging platform with a custom-built phantom. The simulation and experiments were carried out on the single-source CT architecture and the multi-source CT architecture, respectively in the global CT mode and the interior CT mode for comparison. To correct the scattering artifact, two new methods were presented. The first is a beam-stopper-array (BSA)-based method, which enables an online correction of forward scattering and cross scattering simultaneously. The second is a source-trigger-sequence (STS)-based method dedicated to cross-scatter correction. It enables on-the-fly measurements of the cross scattering signals at a few pre-selected views. The CT image quality was quantitatively evaluated in terms of contrast-to-noise ratio (CNR) and CT number deviation before and after the scatter correction.
Results: X-ray cross scattering degraded image quality in both the simulation and experiments. Before the scatter correction, the multi-source interior CT mode yielded a reduction of CNR at the ROIs by up to 68.5% and 50.7% in the simulation and experiments, respectively. The stationary BSA-based method significantly improved CNR and CT number accuracy in the images from multi-source interior CT mode, by reducing the negative effects from both forward scattering and cross scattering. The STS-based method enabled multi-source interior CT mode to provide comparable image quality to that with the single-source interior CT mode, by correcting the artifact from cross scattering. The remaining forward scattering artifact can be corrected with the fast adaptive scatter kernel superposition (FASKS) technique. With the proposed scatter correction methods, the CT number error at the ROIs was reduced to less than 37 HU in both simulation and experiments, respectively.
Conclusions: Cross scattering, in addition to forward scattering, can cause significant image quality degradation in the multi-source interior CT architecture. However, image quality can be significantly improved with the proposed scatter correction methods.
Keywords: cross scattering; forward scattering; interior tomography; multi-source; scatter correction.
© 2016 American Association of Physicists in Medicine.
Similar articles
-
Physics Model-Based Scatter Correction in Multi-Source Interior Computed Tomography.IEEE Trans Med Imaging. 2018 Feb;37(2):349-360. doi: 10.1109/TMI.2017.2741259. Epub 2017 Aug 17. IEEE Trans Med Imaging. 2018. PMID: 28829306
-
Scatter Reduction and Correction for Dual-Source Cone-Beam CT Using Prepatient Grids.Technol Cancer Res Treat. 2016 Jun;15(3):416-27. doi: 10.1177/1533034615587615. Epub 2015 May 24. Technol Cancer Res Treat. 2016. PMID: 26009495 Free PMC article.
-
Strategies for scatter correction in dual source CT.Med Phys. 2010 Nov;37(11):5971-92. doi: 10.1118/1.3504606. Med Phys. 2010. PMID: 21158310
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
A general framework and review of scatter correction methods in x-ray cone-beam computerized tomography. Part 1: Scatter compensation approaches.Med Phys. 2011 Jul;38(7):4296-311. doi: 10.1118/1.3599033. Med Phys. 2011. PMID: 21859031 Review.
Cited by
-
Semi-stationary Multi-source AI-powered Real-time Tomography.IEEE Trans Radiat Plasma Med Sci. 2025 Jan;9(1):118-130. doi: 10.1109/trpms.2024.3433575. Epub 2024 Jul 25. IEEE Trans Radiat Plasma Med Sci. 2025. PMID: 40808799 Free PMC article.
-
Deep Learning-Based Internal Target Volume (ITV) Prediction Using Cone-Beam CT Images in Lung Stereotactic Body Radiotherapy.Technol Cancer Res Treat. 2022 Jan-Dec;21:15330338211073380. doi: 10.1177/15330338211073380. Technol Cancer Res Treat. 2022. PMID: 35188835 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical