. 2017 Mar;38(3):596-602.
doi: 10.3174/ajnr.A5052. Epub 2017 Jan 19.

FRED Flow Diverter: A Study on Safety and Efficacy in a Consecutive Group of 50 Patients

Affiliations

FRED Flow Diverter: A Study on Safety and Efficacy in a Consecutive Group of 50 Patients

H Luecking et al. AJNR Am J Neuroradiol. 2017 Mar.

Abstract

Background and purpose: Endovascular flow diverters are increasingly used for the treatment of cerebral aneurysms. We assessed the safety and efficacy of the Flow-Redirection Endoluminal Device (FRED) in a consecutive series of 50 patients.

Materials and methods: Inclusion criteria were wide-neck, blister-like, or fusiform/dissecting aneurysms independent of size, treated with the FRED between February 2014 and May 2015. Assessment criteria were aneurysm occlusion, manifest ischemic stroke, bleeding, or death. The occlusion rate was assessed at 3 months with flat panel CT and at 6 months with DSA by using the Raymond classification and the O'Kelly-Marotta grading scale.

Results: Fifty patients with 52 aneurysms were treated with 54 FREDs; 20 patients were treated with the FRED and coils. Aneurysm size ranged from 2.0 to 18.5 mm. Deployment of the FRED was successful in all cases. There were no device-associated complications. One patient developed mild stroke symptoms that fully receded within days. There have been no late-term complications so far and no treatment-related mortality. Initial follow-up at 3 months showed complete occlusion in 72.3% of the overall study group, Six-month follow-up showed total and remnant-neck occlusion in 87.2% of patients, distributed over 81.5% of the FRED-only cases and 95.0% of the cases with combined treatment.

Conclusions: The FRED flow diverter is a safe device for the treatment of cerebral aneurysms of various types. Our data reveal high occlusion rates at 3 and 6 months, comparable with those in other flow diverters. Long-term occlusion rates are expected.

PubMed Disclaimer

Figures

Fig 1.
Fig 1.
Sidewall aneurysm of the left vertebral artery (A). Gradual deployment of the FRED device (B and C). Note delayed washout of contrast right after deployment (E) and good visualization of the flow diverter on 2D and 3D imaging (D, F–H) (3D imaging acquired with flat panel CT). D and G, The overall (4 markers at each end) and working length (radiopaque “double helix”) are illustrated.
Fig 2.
Fig 2.
Paraophthalmic aneurysm 6 months after (A) and before (B) implantation of 2 FREDs. Note the “white collar” between the parent vessel and the aneurysm, suggesting endothelialization of the stents.

References

    1. Brinjikji W, Rabinstein AA, Nasr DM, et al. Better outcomes with treatment by coiling relative to clipping of unruptured intracranial aneurysms in the United States, 2001–2008. AJNR Am J Neuroradiol 2011;32:1071–75 10.3174/ajnr.A2453 - DOI - PMC - PubMed
    1. Alshekhlee A, Mehta S, Edgell RC, et al. Hospital mortality and complications of electively clipped or coiled unruptured intracranial aneurysm. Stroke 2010;41:1471–76 10.1161/STROKEAHA.110.580647 - DOI - PubMed
    1. Molyneux A, Kerr R, Stratton I, et al. ; International Subarachnoid Aneurysm Trial (ISAT) Collaborative Group. International Subarachnoid Aneurysm Trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised trial. Lancet 2002;360:1267–74 10.1016/S0140-6736(02)11314-6 - DOI - PubMed
    1. Consoli A, Vignoli C, Renieri L, et al. Assisted coiling of saccular wide-necked unruptured intracranial aneurysms: stent versus balloon. J Neurointerv Surg 2016;8:52–57 10.1136/neurintsurg-2014-011466 - DOI - PubMed
    1. Moret J, Cognard C, Weill A, et al. The “Remodelling Technique” in the treatment of wide neck intracranial aneurysms: angiographic results and clinical follow-up in 56 cases. Interv Neuroradiol 1997;3:21–35 - PubMed

MeSH terms