The influence of length of implant on primary stability: An in vitro study using resonance frequency analysis
- PMID: 28149455
- PMCID: PMC5268121
- DOI: 10.4317/jced.53302
The influence of length of implant on primary stability: An in vitro study using resonance frequency analysis
Abstract
Background: Primary stabilityis not sufficientin less contact area between the implant and bone, the healing process because will be disrupted due to micro-motions and fibrous tissue affects osseointegration.
Material and methods: We implemented an in vitro experimental study of total 135 XiVE® implants were inserted in 22.5 bovine cow ribs with bone quality similar to a type IV human bone. Each rib end received a group of three different implant lengths, which were 8mm, 13mm and 15mm and had the same diameter 3.8mm. Immediately after the implant placement, its primary stability was measured using Osstell Mentor equipment. ANOVA Tukey's honest to test the significant difference were performed for data analysis between the resonance measures of the different lengths of implants. Statistical significance was assessed at a level P< 0.05.
Results: A total of 45 implants were inserted for each length at cortical bone level. A significant difference between the three groups in favor of implant with 15mm length group (P = 0.000).
Conclusions: Increasing dental implant length is considered to play a fundamental role in increasing dental implant primary stability, even in poor bone quality, through controlling the bone preparation process. Key words:Dental implants, primary stability, resonance frequency analysis.
Conflict of interest statement
The authors declare that they have no conflict of interest.
Figures
References
-
- Grassi S, Piattelli A, Ferrari DS, Figueiredo LC, Feres M, Iezzi G. Histologic evaluation of human bone integration on machined and sandblasted acid-etched titanium surfaces in type IV bone. J Oral Implantol. 2007;33:8–12. - PubMed
-
- Farré-Pagés N, Augé-Castro ML, Alaejos-Algarra F, Mareque-Bueno J, Ferrés-Padró E, Hernández-Alfaro F. Relation between bone density and primary implant stability. Med Oral Patol Oral Cir Bucal. 2011;16:e62–7. - PubMed
-
- Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22:743–54. - PubMed
-
- Quesada-García MP, Prados-Sánchez E, Olmedo-Gaya MV, Muñoz-Soto E, González-Rodríguez MP, Vallecillo-Capilla M. Measurement of dental implant stability by resonance frequency analysis: a review of the literature. Med Oral Patol Oral Cir Bucal. 2009;14:e538–46. - PubMed
-
- Boronat López A, Balaguer Martínez J, Lamas Pelayo J, Carrillo García C, Peñarrocha Diago M. Resonance frequency analysis of dental implant stability during the healing period. Med Oral Patol Oral Cir Bucal. 2008;13:e244–7. - PubMed