Accuracy of Optical Coherence Tomography Measurements of Rectus Muscle Insertions in Adult Patients Undergoing Strabismus Surgery
- PMID: 28163119
- DOI: 10.1016/j.ajo.2017.01.025
Accuracy of Optical Coherence Tomography Measurements of Rectus Muscle Insertions in Adult Patients Undergoing Strabismus Surgery
Abstract
Purpose: To assess the accuracy of anterior segment optical coherence tomography (AS-OCT) in measuring the distance of extraocular muscle (EOM) insertion to the limbus to improve preoperative assessment of adult patients undergoing strabismus surgery.
Design: Reliability analysis.
Methods: setting: An institutional practice.
Patient population: Seventy-four adult patients scheduled for strabismus surgery on rectus muscles.
Observation procedure: The distance between the EOM insertion and the limbus was measured preoperatively with AS-OCT. The value was compared with the intraoperative measurement obtained with calipers. Additional measurements included the limbus-anterior chamber angle distance with AS-OCT and the axial length with IOLMaster.
Main outcome measure: Agreement between preoperative AS-OCT and intraoperative measurements. A difference of ≤1 mm was "clinically acceptable."
Results: A total of 144 muscles were analyzed. Thirty-one of 33 reoperated muscles were successfully imaged. AS-OCT measurements were within 1 mm of intraoperative measurements in 77% of all muscles. The accuracy was higher for muscles with no prior surgery (83%), as compared with reoperated muscles (58%). Although the accuracy decreased when comparing reoperations to primary surgeries for the medial (79% to 63%; P = .09; 95% confidence interval [CI], -1.38 to 0.11) and the lateral rectus (81% to 49%; P = .11; 95% CI, -2.06 to 0.22), the difference was not significant. No correlation between limbus-anterior chamber angle distance and axial length was established.
Conclusions: AS-OCT is valuable in identifying EOM insertions in primary strabismus surgeries, but the accuracy decreases in reoperations.
Copyright © 2017 Elsevier Inc. All rights reserved.
Similar articles
-
The accuracy of anterior segment optical coherence tomography (AS-OCT) in localizing extraocular rectus muscles insertions.J AAPOS. 2015 Jun;19(3):233-6. doi: 10.1016/j.jaapos.2015.03.012. J AAPOS. 2015. PMID: 26059668
-
Agreement between intraoperative and anterior segment optical coherence tomography measurements of the extraocular muscle insertion distance in patients with previous strabismus surgery.J AAPOS. 2024 Oct;28(5):104011. doi: 10.1016/j.jaapos.2024.104011. Epub 2024 Sep 19. J AAPOS. 2024. PMID: 39304028
-
Anterior Segment Optical Coherence Tomography of Previously Operated Extraocular Muscles.Am Orthopt J. 2017 Jan;67(1):61-66. doi: 10.3368/aoj.67.1.61. Am Orthopt J. 2017. PMID: 28904216
-
Integrated Intraoperative Optical Coherence Tomography in Strabismus Surgery.Dev Ophthalmol. 2021;61:27-31. doi: 10.1159/000511819. Epub 2021 Feb 16. Dev Ophthalmol. 2021. PMID: 33592620 Review.
-
Reliability of Time Domain AS-OCT in Measuring the Extraocular Rectus Muscle Insertion-Limbus Distances: A Systematic Review and Meta-Analysis.Clin Ophthalmol. 2022 Aug 25;16:2823-2835. doi: 10.2147/OPTH.S381644. eCollection 2022. Clin Ophthalmol. 2022. PMID: 36046574 Free PMC article. Review.
Cited by
-
One-Year Outcome Predictors of Strabismus Surgery from Anterior Segment Optical Coherence Tomography with Multiple B-Scan Averaging.Sci Rep. 2019 Feb 21;9(1):2523. doi: 10.1038/s41598-019-39361-5. Sci Rep. 2019. PMID: 30792463 Free PMC article.
-
Analysis of temporal changes in thickness from conjunctiva to sclera after plication of the medial rectus muscle measured by anterior segment optical coherence tomography.Jpn J Ophthalmol. 2023 Sep;67(5):612-617. doi: 10.1007/s10384-023-01006-6. Epub 2023 Jun 21. Jpn J Ophthalmol. 2023. PMID: 37341849
-
Defining the Clinical Role of Swept-Source Anterior Segment Optical Coherence Tomography in Eyes Undergoing Strabismus Surgery.Clin Ophthalmol. 2022 May 5;16:1449-1465. doi: 10.2147/OPTH.S365622. eCollection 2022. Clin Ophthalmol. 2022. PMID: 35547399 Free PMC article.
-
Outcomes of OCT imaging-based reoperations in strabismus.BMC Ophthalmol. 2025 Jan 21;25(1):34. doi: 10.1186/s12886-024-03816-9. BMC Ophthalmol. 2025. PMID: 39838321 Free PMC article.
-
Changes in conjunctival-scleral thickness after strabismus surgery measured with anterior segment optical coherence tomography.Jpn J Ophthalmol. 2018 Sep;62(5):554-559. doi: 10.1007/s10384-018-0609-2. Epub 2018 Jul 6. Jpn J Ophthalmol. 2018. PMID: 29980870
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials