. 2017 Feb 7;12(2):e0170990.
doi: 10.1371/journal.pone.0170990. eCollection 2017.

Validation of Masimo Pronto 7 and HemoCue 201 for hemoglobin determination in children from 1 to 5 years of age

Affiliations

Validation of Masimo Pronto 7 and HemoCue 201 for hemoglobin determination in children from 1 to 5 years of age

Teresa Shamah Levy et al. PLoS One. .

Abstract

Objective: To evaluate the accuracy and precision of HemoCue 201 (HemoCue) and Masimo Pronto 7 (Masimo) devices for measuring hemoglobin (Hb) in epidemiological studies, having venous blood samples as a gold standard.

Material and methods: We measured Hb concentrations in a field sample of 148 children from one to five years of age. Masimo and HemoCue were used for capillary blood samples and an automatic analyzer for venous blood samples. Regression models with no intercept were constructed to measure precision and predictability, concordance correlations to measure accuracy and precision, and Bland-Altman limits of agreement as well as hierarchical linear models to estimate variance.

Results: Both HemoCue and Masimo underestimated Hb concentrations compared to the gold standard. They respectively yielded the following results: regression coefficients of 0.887 and 0.876 with 98.7% and 98.6% predictability; concordance correlation coefficients of 0.183 (p<0.001) and 0.166 (p<0.001); and Bland-Altman variances of -1.51 and -1.62. With regard to Masimo specifically, the three-level Hierarchical Linear Model showed that 57.9% of total variance stemmed from random errors in repeated measures from the same subject.

Conclusions: HemoCue and Masimo measure lower Hb concentrations than the gold standard. Their accuracy and precision levels are comparable. It is essential to ensure proper use of devices through enhanced training of field workers.

PubMed Disclaimer

Conflict of interest statement

The authors have declared that no competing interests exist.

References

    1. WHO. Global nutrition targets 2025: anaemia policy brief. Geneva: World Health Organization; 2014.
    1. Balarajan Y, Ramakrishnan U, Ozaltin E, Shankar AH, and Subramanian SV. Anaemia in low-income and middle-income countries. Lancet. 2011; 378: 2123–2135. 10.1016/S0140-6736(10)62304-5 - DOI - PubMed
    1. Gutiérrez JP, Rivera-Dommarco J, Shamah-Levy, Villalpando-Hernández S, Franco A, Cuevas-Nasu L,et al. Encuesta Nacional de Salud y Nutrición 2012. Resultados Nacionales. Cuernavaca, México: Instituto Nacional de Salud Pública (MX), 2012.
    1. Morales-Ruán MC, Shamah-Levy T, Mundo-Rosas V, Cuevas-Nasu L, Romero-Martínez M, Villalpando S, et al. Food assistance programs in Mexico, coverage and targeting. Salud Publica Mex 2013;55 (2):199–205. - PubMed
    1. WHO. Haemoglobin concentrations for the diagnosis of anaemia and assessment of severity Vitamin and Mineral Nutrition Information System. Geneva, World Health Organization, 2011. (WHO/NMH/NHD/MNM/11.1) (http://www.who.int/vmnis/indicators/haemoglobin.pdf, accessed [feb 9, 2016].

LinkOut - more resources