Smartphone assessment of knee flexion compared to radiographic standards
- PMID: 28179062
- PMCID: PMC5359022
- DOI: 10.1016/j.knee.2016.11.014
Smartphone assessment of knee flexion compared to radiographic standards
Abstract
Background: Measuring knee range of motion (ROM) is an important assessment for the outcomes of total knee arthroplasty. Recent technological advances have led to the development and use of accelerometer-based smartphone applications to measure knee ROM. The purpose of this study was to develop, standardize, and validate methods of utilizing smartphone accelerometer technology compared to radiographic standards, visual estimation, and goniometric evaluation.
Methods: Participants used visual estimation, a long-arm goniometer, and a smartphone accelerometer to determine range of motion of a cadaveric lower extremity; these results were compared to radiographs taken at the same angles.
Results: The optimal smartphone position was determined to be on top of the leg at the distal femur and proximal tibia location. Between methods, it was found that the smartphone and goniometer were comparably reliable in measuring knee flexion (ICC=0.94; 95% CI: 0.91-0.96). Visual estimation was found to be the least reliable method of measurement.
Conclusions: The results suggested that the smartphone accelerometer was non-inferior when compared to the other measurement techniques, demonstrated similar deviations from radiographic standards, and did not appear to be influenced by the person performing the measurements or the girth of the extremity.
Keywords: Accelerometer; Measurement; Motion; Smartphone.
Copyright © 2016 Elsevier B.V. All rights reserved.
Figures
References
-
- Edwards JZ, Greene KA, Davis RS, Kovacik MW, Noe DA, Askew MJ. Measuring flexion in knee arthroplasty patients. J Arthroplasty. 2004;19:369–72. - PubMed
-
- Ferriero G, Vercelli S, Sartorio F, Munoz Lasa S, Ilieva E, Brigatti E, et al. Reliability of a smartphone-based goniometer for knee joint goniometry. Int J Rehabil Res. 2013;36:146–51. - PubMed
-
- Bedekar N, Suryawanshi M, Rairikar S, Sancheti P, Shyam A. Inter and intra-rater reliability of mobile device goniometer in measuring lumbar flexion range of motion. J Back Musculoskelet Rehabil. 2013 - PubMed
-
- Ockendon M, Gilbert RE. Validation of a novel smartphone accelerometer-based knee goniometer. J Knee Surg. 2012;25:341–5. - PubMed
-
- Box G, Hunter W, Hunter J. Statisics for Experimenters: An introduction to Design, Data Analysis, and Model Building. Toronto, Ontario, Canada: John Wiley & Sons; 1978.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources