Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 Jan;47(1):1-4.
doi: 10.4274/tjo.92593. Epub 2017 Jan 17.

Comparison of Intraocular Pressure Measurements in Healthy Pediatric Patients using Three Types of Tonometers

Affiliations

Comparison of Intraocular Pressure Measurements in Healthy Pediatric Patients using Three Types of Tonometers

Muhsin Eraslan et al. Turk J Ophthalmol. 2017 Jan.

Abstract

Objectives: This study aimed to compare intraocular pressure (IOP) measurements in healthy pediatric patients using three types of tonometers.

Materials and methods: Seventy-eight eyes of 78 patients under the age of 18 who underwent a routine ophthalmologic examination were included in the study. IOP was measured using Tono-Pen (TP) tonometry, Goldmann applanation tonometry (GAT), and non-contact tonometry (NCT), consecutively. IOP was adjusted based on central corneal thickness (CCT). Patients with any ocular disorders other than a limited refractive error were excluded from the study.

Results: The study consisted of 46 girls and 32 boys. The mean age was 12.6±2.7 (range: 5-17) years. The mean CCT was 559.3±35.3 µm. The mean refractive error was -0.50±1.70. The mean level of visual acuity was 0.98±0.1 (range: 0.3-1.0) using the Snellen chart. Significant differences were found between the measurement results of each of the three tonometric methods. Mean IOP was 12.1±2.2 mmHg for TP, 15.7±2.5 mmHg for GAT, and 17.1±3.1 mmHg for NCT. The correlations between measurement methods revealed that the highest correlation was between NCT and GAT (p<0.001, r=0.670). The second highest correlation was between NCT and TP (p<0.001, r=0.477). The lowest correlation was between GAT and TP (p<0.001, r=0.403). A positive correlation was found between CCT and each IOP measurement method.

Conclusion: In pediatric patients, TP and NCT measurements were found to be positively correlated with GAT measurements. Because TP measurements were lower than GAT measurements and NCT measurements were higher than GAT measurements, patient follow-ups, treatment strategies, and surgery plans must be organized taking these differences into consideration.

Keywords: Goldmann applanation tonometry; Pediatric intraocular pressure; Tono-Pen tonometry; central corneal thickness; non-contact tonometry.

PubMed Disclaimer

Conflict of interest statement

No conflict of interest was declared by the authors. Financial Disclosure: The authors declared that this study received no financial support.

Figures

Figure 1
Figure 1. Bland-Altman scatter plot showing the errors in intraocular pressure measurements obtained with the Tono-Pen compared to Goldmann applanation tonometry results TP: Tono-Pen, GAT: Goldmann applanation tonometry
Figure 2
Figure 2. Bland-Altman scatter plot showing the errors in intraocular pressure measurements obtained with non-contact tonometer compared to Goldmann applanation tonometry results NCT: Non-contact tonometer, GAT: Goldmann applanation tonometry
Figure 3
Figure 3. Correlation between the results of Goldmann applanation tonometry and measurements obtained using the Tono-Pen and non-contact tonometer. A moderately significant correlation was found between the measurements TP: Tono-Pen, NCT: Non-contact tonometer

References

    1. Kass MA, Heuer DK, Higginbotham EJ, Johnson CA, Keltner JL, Miller JP, KParrish RK, Wilson MR, Gordon MO. The Ocular Hypertension Treatment Study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120:701–713. - PubMed
    1. Kao SF, Lichter PR, Bergstrom TJ, Rowe S, Musch DC. Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. Ophthalmology. 1987;94:1541–1544. - PubMed
    1. Patel H, Gilmartin B, Cubbidge RP, Logan NS. In vivo measurement of regional variation in anterior scleral resistance to Schiotz indentation. Ophthalmic Physiol Opt. 2011;31:437–443. - PubMed
    1. Gupta S, Sinha G, Sharma R, Nayak B, Patil B, Kashyap B, Shameer A, Dada T. Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication. Int Ophthalmol. 2016;36:9–15. - PubMed
    1. Shousha SM, Abo Steit MA, Hosny MH, Ewais WA, Shalaby AM. Comparison of different intraocular pressure measurement techniques in normal eyes, post surface and post lamellar refractive surgery. Clin Ophthalmol. 2013;7:71–79. - PMC - PubMed

LinkOut - more resources