Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Multicenter Study
. 2017 Feb 13;10(3):276-282.
doi: 10.1016/j.jcin.2016.11.025.

Initial Experience of a Second-Generation Self-Expanding Transcatheter Aortic Valve: The UK & Ireland Evolut R Implanters' Registry

Affiliations
Free article
Multicenter Study

Initial Experience of a Second-Generation Self-Expanding Transcatheter Aortic Valve: The UK & Ireland Evolut R Implanters' Registry

Sundeep S Kalra et al. JACC Cardiovasc Interv. .
Free article

Abstract

Objectives: The authors present the UK and Irish real-world learning curve experience of the Evolut R transcatheter heart valve.

Background: The Evolut R is a self-expanding, repositionable, and fully recapturable second-generation transcatheter heart valve with several novel design features to improve outcomes and reduce complications.

Methods: Clinical, procedural, and 30-day outcome data were prospectively collected for the first 264 patients to receive the Evolut R valve in the United Kingdom and Ireland.

Results: A total of 264 consecutive Evolut R implantations were performed across 9 centers. The mean age was 81.1 ± 7.8 years, and the mean logistic European System for Cardiac Operative Risk Evaluation score was 19.9 ± 13.7%. Procedural indications included aortic stenosis (72.0%), mixed aortic valve disease (17.4%), and failing aortic valve bioprostheses (10.6%). Conscious sedation was used in 39.8% of patients and transfemoral access in 93.6%. The procedural success rate was 91.3%, and paravalvular leak immediately after implantation was mild or less in 92.3%. Major complications were rare: cardiac tamponade in 0.4%, conversion to sternotomy in 0.8%, annular rupture in 0.0%, coronary occlusion in 0.8%, major vascular in 5.3%, acute kidney injury in 6.1%, new permanent pacemaker implantation in 14.7%, and procedure-related death in 0.0%. At 30-day follow-up, survival was 97.7%, paravalvular leak was mild or less in 92.3%, and the stroke rate was 3.8%.

Conclusions: This registry represents the largest published real-world experience of the Evolut R valve. The procedural success rate was high and safety was excellent, comparable with previous studies of the Evolut R valve and other second-generation devices. The low rate of complications represents an improvement on first-generation devices.

Keywords: Evolut R; aortic stenosis; complication; outcome; transcatheter aortic valve replacement; transcatheter heart valve.