Excellent mid-term results of a new polished tapered modular cemented stem: a study of 113 hip replacements with minimum 5-year follow-up
- PMID: 28222212
- PMCID: PMC6380089
- DOI: 10.5301/hipint.5000486
Excellent mid-term results of a new polished tapered modular cemented stem: a study of 113 hip replacements with minimum 5-year follow-up
Abstract
Background: This prospective cohort study reports the medium-term clinical and radiographic results of 113 hips treated with a hybrid total hip replacement (THR) with a new cemented tapered cobalt-chrome (Co-Cr) stem with a titanium (Ti) modular neck (ProfemurXm®).
Method: Between October 2008 and December 2010 we performed 115 consecutive hybrid THR with the ProfemurXm® in 105 patients.
Results: Survivorship of the implant (stem and modular neck) at a mean of 6.5 years (min 5-max 8) was 100% with the endpoint revision for any reason. No implant was at risk for revision or showed signs of loosening. The mean Harris Hip Score was 89/100, mean Oxford Hip Score was 43/48, mean WOMAC was 91/100. No patient had thigh pain, no patient reported squeaking. There were no dislocations in this cohort. No implant showed development of radiolucent lines (RLL), either at the stem-cement or cement-bone interface. No hip showed osteolysis or calcar resorption. The mean femoral subsidence of the stem within the cement mantel was 0.31 mm (range 0-0.6 mm) after 6.5 years. With the use of this modular stem, 93% of hips showed no measurable leg length difference after THR, and leg length could be restored within a 5-mm limit in 99% of hips.
Conclusions: The mid-term results of this new polished stem were excellent, without adverse effects from the use of modularity.
Conflict of interest statement
Conflict of interest: None.
References
-
- Ling RS Charity J Lee AJ Whitehouse SL Timperley AJ Gie GA The long-term results of the original Exeter polished cemented femoral component: a follow-up report. J Arthroplasty 2009;24(4):511–517. - PubMed
-
- Simpson DJ Little JP Gray H Murray DW Gill HS Effect of modular neck variation on bone and cement mantle mechanics around a total hip arthroplasty stem. Clin Biomech (Bristol, Avon) 2009;24(3):274–285. - PubMed
-
- Brooker AF Bowerman JW Robinson RA Riley LH Jr Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 1973;55(8):1629–1632. - PubMed