. 2017 May;37(3):240-247.
doi: 10.3343/alm.2017.37.3.240.

Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer

Affiliations

Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer

In Young Yoo et al. Ann Lab Med. 2017 May.

Abstract

Background: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluorescent ANA (FANA) testing. We evaluated the performance of EUROPattern Suite (Euroimmun AG, Germany), an automated FANA image analyzer, with regard to ANA detection and pattern recognition compared with conventional manual interpretation using the fluorescence microscopic IIF assay.

Methods: A total of 104 samples including 70 ANA-positive sera and 34 ANA-negative sera collected from September to October 2015 were included. The sensitivity, specificity, and pattern recognition function were evaluated to determine the performance of EUROPattern Suite compared with the manual IIF assay results.

Results: The sensitivity and specificity of EUROPattern Suite for ANA detection were 94.3% and 94.1%, respectively. The concordance rate between the two methods was 94.2%. For pattern recognition, 45.7% of the samples were assigned identical ANA patterns including simple and mixed. When major pattern matching was considered, 83.7% (41/49) and 95.2% (20/21) of the samples with simple and mixed patterns, respectively, showed concordant results between the two methods.

Conclusions: EUROPattern Suite, an automated FANA image analyzer, provides a viable option for distinguishing between positive and negative results, although the ability to assign specific patterns is insufficient to replace manual microscopic interpretation. This automated system may increase efficiency in laboratories, in which a large number of samples need to be processed.

Keywords: Antinuclear antibody; Automated image analyzer; Indirect immunofluorescence assay; Pattern recognition.

PubMed Disclaimer

Conflict of interest statement

Authors' Disclosures of Potential Conflicts of Interest: No potential conflicts of interest relevant to this article were reported.

Figures

Fig. 1
Fig. 1. Images of mixed patterns that were assigned discordant patterns by EUROPattern Suite compared with manual microscopic interpretation. (A) Homogenous/Centromere, (B) Homogenous/Nuclear dot, (C) Homogenous/Nuclear membrane, (D) Centromere/Nuclear membrane/Cytoplasmic with mitochondrial, (E) Nucleolar/Nuclear dot, and (F) Speckled/Nucleolar. Each image matches the numbers in Table 4 as follows: #10, #11, #13, #17, #19, and #21, respectively.

References

    1. Voigt J, Krause C, Rohwäder E, Saschenbrecker S, Hahn M, Danckwardt M, et al. Automated indirect immunofluorescence evaluation of antinuclear autoantibodies on HEp-2 cells. Clin Dev Immunol. 2012;2012:651058. - PMC - PubMed
    1. Meroni PL, Schur PH. ANA screening: an old test with new recommendations. Ann Rheum Dis. 2010;69:1420–1422. - PubMed
    1. Roggenbuck D, Hiemann R, Bogdanos D, Reinhold D, Conrad K. Standardization of automated interpretation of immunofluorescence tests. Clin Chim Acta. 2013;421:168–169. - PubMed
    1. Bizzaro N, Antico A, Platzgummer S, Tonutti E, Bassetti D, Pesente F, et al. Automated antinuclear immunofluorescence antibody screening: a comparative study of six computer-aided diagnostic systems. Autoimmun Rev. 2014;13:292–298. - PubMed
    1. Schouwers S, Bonnet M, Verschueren P, Westhovens R, Blockmans D, Marien G, et al. Value-added reporting of antinuclear antibody testing by automated indirect immunofluorescence analysis. Clin Chem Lab Med. 2014;52:547–551. - PubMed

Publication types

MeSH terms

Substances

LinkOut - more resources