Validation of a hand-held point of care device for lactate in adult and pediatric patients using traditional and locally-smoothed median and maximum absolute difference curves
- PMID: 28235428
- DOI: 10.1016/j.cca.2017.02.015
Validation of a hand-held point of care device for lactate in adult and pediatric patients using traditional and locally-smoothed median and maximum absolute difference curves
Abstract
Background: Lactate is commonly used in septic patients and is a viable biomarker for trauma patients. Its pre-hospital use could assist triaging and managing patients with these conditions.
Methods: We evaluated the analytical performance of the point-of-care (POC) StatStrip Xpress Lactate Meter (Nova Biomedical) and compared it to the ABL 800 (Radiometer). We measured lactate in 250 adult and 250 pediatric whole blood samples in 2 laboratories. The performance of the POC meter was assessed by traditional linear regression and Bland-Altman plots, and locally-smoothed (LS) median absolute difference and maximum absolute difference (MAD and MaxAD) curves.
Results: The StatStrip was linear with acceptable reproducibility at clinically relevant concentrations. Correlation with the ABL800 showed a negative bias for both populations with slope, bias ±SD (% bias) of 0.78, -0.4±0.7 (-14.5%) in children and 0.80-0.3±0.6 (-13.3%) in adults. The proportional bias appeared more significant at concentrations >4mmol/l (36.0mg/dl). The StatStrip misclassified 7.6 and 8.8% pediatric and adult samples, respectively, to lower risk categories defined using guidelines driven cut-offs. The LS MAD curves identified one breakout, concentration where the LS MAD exceeds the total allowable error limit of 0.3mmol/l (2.7mg/dl), at lactate concentrations of 3.8 and 3.2mmol/l (34.2 and 28.8mg/dl) in the pediatric and adult curves, respectively. Breakthroughs, points at which the LS MaxAD curve exceeds the 95th percentile of MaxADs, occur at concentrations above 7.5mmol/l (67.6mg/dl) for both populations where the performance of the POC meter became erratic. We concluded that if serial lactate measurements are performed, the same method should be used for baseline and follow up measurements. The LS MAD and LS MaxAD curves allowed visual and quantitative mapping of the performance of the lactate POC meter over the range of concentrations measured.
Conclusions: This approach seems useful for the identification of points at which the performance of a POC meter differs significantly from a comparison method and thresholds of poor analytical performance.
Keywords: Lactate; Locally-smoothed maximum absolute difference curve; Locally-smoothed median absolute difference curve; Point-of-care; Pre-hospital; Sepsis; Trauma; Validation.
Copyright © 2017 Elsevier B.V. All rights reserved.
Similar articles
-
Mapping point-of-care performance using locally-smoothed median and maximum absolute difference curves.Clin Chem Lab Med. 2011 Oct;49(10):1637-46. doi: 10.1515/CCLM.2011.655. Clin Chem Lab Med. 2011. PMID: 22008003
-
Analytical performance of three whole blood point-of-care lactate devices compared to plasma lactate comparison methods and a flow-injection mass spectrometry method.Clin Biochem. 2017 Mar;50(4-5):168-173. doi: 10.1016/j.clinbiochem.2016.11.004. Epub 2016 Nov 9. Clin Biochem. 2017. PMID: 27836621
-
Validation of a point-of-care (POC) lactate testing device for fetal scalp blood sampling during labor: clinical considerations, practicalities and realities.Clin Chem Lab Med. 2014 Jun;52(6):825-33. doi: 10.1515/cclm-2013-0732. Clin Chem Lab Med. 2014. PMID: 24406288
-
Point-of-care devices for physiological measurements in field conditions. A smorgasbord of instruments and validation procedures.Comp Biochem Physiol A Mol Integr Physiol. 2016 Dec;202:99-111. doi: 10.1016/j.cbpa.2016.04.009. Epub 2016 Apr 12. Comp Biochem Physiol A Mol Integr Physiol. 2016. PMID: 27083239 Review.
-
Diagnostic Use of Lactate in Exotic Animals.Vet Clin North Am Exot Anim Pract. 2022 Sep;25(3):585-596. doi: 10.1016/j.cvex.2022.05.006. Vet Clin North Am Exot Anim Pract. 2022. PMID: 36122941 Review.
Cited by
-
Accuracy of a point-of-care blood lactate measurement device in a prehospital setting.J Clin Monit Comput. 2022 Dec;36(6):1679-1687. doi: 10.1007/s10877-022-00812-6. Epub 2022 Jan 27. J Clin Monit Comput. 2022. PMID: 35084641 Free PMC article.
-
Added predictive value of prehospital measurement of point-of-care lactate in an adult general EMS population in Sweden: a multi-centre observational study.Scand J Trauma Resusc Emerg Med. 2024 Aug 20;32(1):72. doi: 10.1186/s13049-024-01245-7. Scand J Trauma Resusc Emerg Med. 2024. PMID: 39164765 Free PMC article.
-
Prehospital lactate levels in blood as a seizure biomarker: A multi-center observational study.Epilepsia. 2021 Feb;62(2):408-415. doi: 10.1111/epi.16806. Epub 2021 Jan 8. Epilepsia. 2021. PMID: 33417237 Free PMC article.
-
Determinants of prehospital lactate in trauma patients: a retrospective cohort study.BMC Emerg Med. 2020 Mar 11;20(1):18. doi: 10.1186/s12873-020-00314-1. BMC Emerg Med. 2020. PMID: 32160880 Free PMC article.
-
POCT: An Inherently Ideal Tool in Pediatric Laboratory Medicine.EJIFCC. 2021 Jun 29;32(2):145-157. eCollection 2021 Jun. EJIFCC. 2021. PMID: 34421483 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous