Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 Feb 17;5(1):e000320.
doi: 10.1136/bmjdrc-2016-000320. eCollection 2017.

Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus

Affiliations

Performance of the FreeStyle Libre Flash glucose monitoring system in patients with type 1 and 2 diabetes mellitus

M J Fokkert et al. BMJ Open Diabetes Res Care. .

Abstract

Objective: To evaluate the performance of the FreeStyle Libre Flash continuous glucose monitoring (FSL-CGM) system against established central laboratory methods.

Research design and methods: 20 subjects (8 type 1 diabetes mellitus, 12 type 2 diabetes mellitus) were analyzed. FSL-CGM sensor measurements (inserted in arm and abdomen) were compared with capillary blood glucose results analyzed with StatStrip as semigold standard. The glucose response after a standardized oral glucose load was measured by FSL-CGM and capillary samples analyzed by perchloric acid hexokinase (PCA-HK) method, StatStrip and FSL test strip (FSLC), and a commonly used CGM system (iPro2).

Results: FSL-CGM arm sensor readings showed 85.5% of paired readings falling within Clarke Error Grid (ISO 15197:2013) zone A when compared with StatStrip. For FSL-CGM abdomen and FSLC, these percentages were 64% and 98%, respectively. The overall correlation of FSL-CGM in the arm and the StatStrip indicates a performance with lower results with the FSL-CGM in the arm than expected based on the StatStrip in the lower glucose ranges, and higher results than expected in the higher ranges. Following a standardized glucose load, a slower rise in glucose level was observed for FSL-CGM arm as compared with PCA-HK, StatStrip, FSLC, and iPro2 during the first 45-60 min after glucose load ingestion.

Conclusions: Certain matters need attention while using the FSL-CGM in daily life including the observed lower values in the lower ranges, and the underestimation of the effect of a meal on glucose response. These effects of such deviations can partly be overcome by optimizing the available user instructions.

Trial registration number: TC5348; results.

Keywords: Blood Glucose Self-Monitoring; Continuous Glucose Monitoring; Endocrinology Diabetes; Glucose Control.

PubMed Disclaimer

Conflict of interest statement

Competing interests: None declared.

Figures

Figure 1
Figure 1
Clarke Error Grids (ISO15197; 2013) of StatStrip and coincident FSLTab-CGM arm (upper left panel), FSL-CGM abdomen (lower left panel) and FSLC readings (right panel) in the combined population of type 1 and 2 diabetes. Number and percentages of paired readings within error grid zones A, B, C D and E are presented. FSL-CGM, FreeStyle Libre Flash continuous glucose monitoring.
Figure 2
Figure 2
The 75 g glucose loading test. Comparison of FSL-CGM versus PCA-HK, StatStrip, FSLC test strip, and iPro2; and PCA-HK versus FSLC test strip, and iPro2. Glucose concentrations (mg/dL; median, interquartile Q1 and Q3) at time points 0, 15, 30, 45, 60, 75 and 90 min in the combined population of type 1 and 2 diabetes. *p Value ≤0.05; **p Value ≤0.01. FSL-CGM, FreeStyle Libre Flash continuous glucose monitoring; PCA-HK, perchloric acid hexokinase.

References

    1. ISO 15197. In vitro diagnostic test systems—requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus 2013. https://http://www.iso.org/obp/ui (accessed 29 Feb 2016).
    1. Rodbard D. Continuous glucose monitoring: a review of successes, challenges, and opportunities. Diabetes Technol Ther 2016;18(Suppl 2):S23–S213. 10.1089/dia.2015.0417 - DOI - PMC - PubMed
    1. Langendam M, Luijf YM, Hooft L et al. . Continuous glucose monitoring systems for type 1 diabetes mellitus. Cochrane Database Syst Rev 2012;(1):CD008101 10.1002/14651858.CD008101.pub2 - DOI - PMC - PubMed
    1. Poolsup N, Suksomboon N, Kyaw AM. Systematic review and meta-analysis of the effectiveness of continuous glucose monitoring (CGM) on glucose control in diabetes. Diabetol Metab Syndr 2013;5:39 10.1186/1758-5996-5-39 - DOI - PMC - PubMed
    1. Riemsma R, Corro Ramos I, Birnie R et al. . Integrated sensor-augmented pump therapy systems [the MiniMed(R) Paradigm Veo system and the Vibe and G4(R) PLATINUM CGM (continuous glucose monitoring) system] for managing blood glucose levels in type 1 diabetes: a systematic review and economic evaluation. Health Technol Assess 2016;20:1–252. - PMC - PubMed

LinkOut - more resources