Two-piece ALIF cage optimizes the bone-implant interface in a 360° setting
- PMID: 28247079
- DOI: 10.1007/s00586-017-5009-7
Two-piece ALIF cage optimizes the bone-implant interface in a 360° setting
Abstract
Introduction: Clinical research has documented that cage subsidence and the loss of balance correction is a significant complication related to spinal fusion. Subsidence is a multifactorial complication, where implant design is one important element. The aim of the study is to compare the rigidity and bone-implant relative motion of segments treated with either a conventional one-piece ALIF cage versus a two-piece ALIF cage, which adapts in situ and permits 7°-21° of lordosis.
Methods: Seven lumbosacral (L3-S1) human cadaver specimens were tested in a universal spine tester in the intact condition, the specimens instrumented with a two-piece ALIF cage (Statur®-L, FBC Device, Denmark) and a one-piece ALIF cage (Pezo™-A, Ulrich GmbH & Co. KG., Germany), both supplemented with a pedicle screw system using pure moments of ±7.5 Nm in three principal motion directions. For assessment of the bone-implant interface, fluoroscopic videos were captured during motion and 3D motion was measured using an optical motion capturing system.
Results: Significantly less motion at the implant-endplate interface was found for the two-piece cage (1.0° ± 0.6°) in comparison to the one-piece cage (4.2° ± 1.7°) in flexion/extension. No significant differences in segment rigidity were found between the one-piece and two-piece cages in the 360° setup, while both configurations significantly reduced the range of motion compared to the intact condition (p < 0.05).
Conclusion: In comparison to the traditional one-piece ALIF cages, the two-piece cage concept reduced the relative motion at the bone-implant interface without compromising stability.
Keywords: ALIF; Biomechanics; In vitro; Lumbar spine; Sagittal balance.
Similar articles
-
Biomechanical assessment of anterior lumbar interbody fusion with an anterior lumbosacral fixation screw-plate: comparison to stand-alone anterior lumbar interbody fusion and anterior lumbar interbody fusion with pedicle screws in an unstable human cadaver model.Spine (Phila Pa 1976). 2006 Apr 1;31(7):762-8. doi: 10.1097/01.brs.0000206360.83728.d2. Spine (Phila Pa 1976). 2006. PMID: 16582849
-
In vitro study of biomechanical behavior of anterior and transforaminal lumbar interbody instrumentation techniques.Neurosurgery. 2006 Dec;59(6):1271-6; discussion 1276-7. doi: 10.1227/01.NEU.0000245609.01732.E4. Neurosurgery. 2006. PMID: 17277690
-
[Mid-term results of 360-degree lumbar spondylodesis with the use of a tantalum implant for disc replacement].Acta Chir Orthop Traumatol Cech. 2009 Oct;76(5):388-93. Acta Chir Orthop Traumatol Cech. 2009. PMID: 19912702 Czech.
-
Influence of cage geometry on sagittal alignment in instrumented posterior lumbar interbody fusion.Spine (Phila Pa 1976). 2003 Aug 1;28(15):1693-9. doi: 10.1097/01.BRS.0000083167.78853.D5. Spine (Phila Pa 1976). 2003. PMID: 12897494
-
MIS Expandable Interbody Spacers: A Literature Review and Biomechanical Comparison of an Expandable MIS TLIF With Conventional TLIF and ALIF.Spine (Phila Pa 1976). 2016 Apr;41 Suppl 8:S44-9. doi: 10.1097/BRS.0000000000001465. Spine (Phila Pa 1976). 2016. PMID: 26825792 Review.
Cited by
-
Endplate Trauma During Implant Insertion Affects the Expulsion Risk of Anterior Lumbar Interbody Fusion Devices.Cureus. 2023 Mar 29;15(3):e36845. doi: 10.7759/cureus.36845. eCollection 2023 Mar. Cureus. 2023. PMID: 37123763 Free PMC article.
-
Range of Motion and Neutral Zone of All Human Spinal Motion Segments: A Data Collection of 30 Years of In Vitro Experiments Performed Under Standardized Testing Conditions.JOR Spine. 2025 Mar 5;8(1):e70052. doi: 10.1002/jsp2.70052. eCollection 2025 Mar. JOR Spine. 2025. PMID: 40046266 Free PMC article.
-
Augmenting Screw Technique to Prevent TLIF Cage Subsidence: A Biomechanical In Vitro Study.Bioengineering (Basel). 2025 Mar 24;12(4):337. doi: 10.3390/bioengineering12040337. Bioengineering (Basel). 2025. PMID: 40281697 Free PMC article.
-
Evaluation of a Force-Control Experimental Method to Perform Unconstrained Load-Induced Subsidence Testing of Spinal Interbody Implants.JOR Spine. 2025 Jul 25;8(3):e70087. doi: 10.1002/jsp2.70087. eCollection 2025 Sep. JOR Spine. 2025. PMID: 40718821 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources