Performance of the SAMBA I and II HIV-1 Semi-Q Tests for viral load monitoring at the point-of-care

Affiliations

Performance of the SAMBA I and II HIV-1 Semi-Q Tests for viral load monitoring at the point-of-care

Neha Goel et al. J Virol Methods. 2017 Jun.

Abstract

Although access to antiretroviral therapy for HIV infection is increasing in resource-poor countries, viral load testing for monitoring of treatment efficacy remains limited, expensive, and confined to centralized laboratories. The SAMBA HIV-1 Semi-Q Test is a nucleic acid-based amplification assay developed for viral load monitoring performed on either the semi-automated SAMBA I system for laboratory use or the fully automated SAMBA II system for point-of care use. We have assessed the performance characteristics of the SAMBA HIV-1 Semi-Q Test on SAMBA I and SAMBA II systems according to the Common Technical Specifications of the European Community's 98/79 In Vitro Diagnostic Medical Devices Directive. The sensitivity, specificity, reproducibility, and viral subtype coverage of the test were similar on the SAMBA I and SAMBA II platforms. The clinical performance on the SAMBA I system was compared with the Roche CAP/CTM assay and evaluated in-house with 130 patient specimens from London as well as in the field with 390 specimens in Kenya and Zimbabwe. The overall concordance between the SAMBA and CAP/CTM assays was 98.1%. The clinical performance of the test on the SAMBA II platform in comparison with the Abbott HIV-1 RealTime Assay was evaluated in-house with 150 specimens from Ukraine, yielding a concordance of 98.0%. The results thus show that the SAMBA HIV-1 Semi-Q Test performs equivalently on SAMBA I and SAMBA II, and they suggest that the test is suitable for implementation at the point-of-care in resource-poor regions where viral load testing is desperately needed but often unavailable.

Keywords: Diagnosis; HIV-1; Point-of-care; Therapy monitoring; Viral load.