Flow and wall shear stress characterization after endovascular aneurysm repair and endovascular aneurysm sealing in an infrarenal aneurysm model
- PMID: 28285931
- DOI: 10.1016/j.jvs.2016.10.077
Flow and wall shear stress characterization after endovascular aneurysm repair and endovascular aneurysm sealing in an infrarenal aneurysm model
Abstract
Background: Endovascular aneurysm repair (EVAR) with a modular endograft has become the preferred treatment for abdominal aortic aneurysms. A novel concept is endovascular aneurysm sealing (EVAS), consisting of dual endoframes surrounded by polymer-filled endobags. This dual-lumen configuration is different from a bifurcation with a tapered trajectory of the flow lumen into the two limbs and may induce unfavorable flow conditions. These include low and oscillatory wall shear stress (WSS), linked to atherosclerosis, and high shear rates that may result in thrombosis. An in vitro study was performed to assess the impact of EVAR and EVAS on flow patterns and WSS.
Methods: Four abdominal aortic aneurysm phantoms were constructed, including three stented models, to study the influence of the flow divider on flow (Endurant [Medtronic, Minneapolis, Minn], AFX [Endologix, Irvine, Calif], and Nellix [Endologix]). Experimental models were tested under physiologic resting conditions, and flow was visualized with laser particle imaging velocimetry, quantified by shear rate, WSS, and oscillatory shear index (OSI) in the suprarenal aorta, renal artery (RA), and common iliac artery.
Results: WSS and OSI were comparable for all models in the suprarenal aorta. The RA flow profile in the EVAR models was comparable to the control, but a region of lower WSS was observed on the caudal wall compared with the control. The EVAS model showed a stronger jet flow with a higher shear rate in some regions compared with the other models. Small regions of low WSS and high OSI were found near the distal end of all stents in the common iliac artery compared with the control. Maximum shear rates in each region of interest were well below the pathologic threshold for acute thrombosis.
Conclusions: The different stent designs do not influence suprarenal flow. Lower WSS is observed in the caudal wall of the RA after EVAR and a higher shear rate after EVAS. All stented models have a small region of low WSS and high OSI near the distal outflow of the stents.
Copyright © 2016 Society for Vascular Surgery. Published by Elsevier Inc. All rights reserved.
Similar articles
-
The Influence of Positioning of the Nellix Endovascular Aneurysm Sealing System on Suprarenal and Renal Flow: An In Vitro Study.J Endovasc Ther. 2017 Oct;24(5):677-687. doi: 10.1177/1526602817719465. Epub 2017 Jul 10. J Endovasc Ther. 2017. PMID: 28689483
-
Partial renal coverage in endovascular aneurysm repair causes unfavorable renal flow patterns in an infrarenal aneurysm model.J Vasc Surg. 2018 May;67(5):1585-1594. doi: 10.1016/j.jvs.2017.05.092. Epub 2017 Oct 12. J Vasc Surg. 2018. PMID: 28893490
-
Effects on Aortoiliac Fluid Dynamics After Endovascular Sealing of Abdominal Aneurysms.Vasc Endovascular Surg. 2018 Nov;52(8):621-628. doi: 10.1177/1538574418791059. Epub 2018 Jul 29. Vasc Endovascular Surg. 2018. PMID: 30058480
-
Studying the interaction of stent-grafts and treated abdominal aortic aneurysms: time to move caudally!J Endovasc Ther. 2015 Jun;22(3):413-20. doi: 10.1177/1526602815583494. J Endovasc Ther. 2015. PMID: 25991770 Review.
-
Polymer leak with the Ovation Abdominal Stent Graft System: Early recognition and treatment.Vascular. 2020 Apr;28(2):159-164. doi: 10.1177/1708538119862923. Epub 2019 Oct 9. Vascular. 2020. PMID: 31597539 Review. No abstract available.
Cited by
-
Experimental Analysis of Pressure and Flow Alterations During and After Insertion of a Multilayer Flow Modulator into an AAA Model with Incorporated Branch.Cardiovasc Intervent Radiol. 2021 Aug;44(8):1251-1259. doi: 10.1007/s00270-021-02835-z. Epub 2021 Apr 27. Cardiovasc Intervent Radiol. 2021. PMID: 33907900
-
Editorial: Clinical and Hemodynamic Performance of Aortic Endografts.Front Surg. 2018 Jun 22;5:44. doi: 10.3389/fsurg.2018.00044. eCollection 2018. Front Surg. 2018. PMID: 29988320 Free PMC article. No abstract available.
-
Endovascular Aneurysm Sealing is Associated with Higher Medium-Term Survival than Traditional EVAR.Ann Vasc Surg. 2020 Feb;63:145-154. doi: 10.1016/j.avsg.2019.08.094. Epub 2019 Oct 16. Ann Vasc Surg. 2020. PMID: 31629124 Free PMC article.
-
A novel roller pump for physiological flow.Artif Organs. 2020 Aug;44(8):818-826. doi: 10.1111/aor.13670. Epub 2020 Mar 25. Artif Organs. 2020. PMID: 32065666 Free PMC article.
-
Long-Term Remodeling of Aortoiliac Vessels After Standard EVAR, the Reality to Be Considered.J Clin Med. 2025 Aug 8;14(16):5626. doi: 10.3390/jcm14165626. J Clin Med. 2025. PMID: 40869452 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials