Comparison of spin-echo echoplanar imaging and gradient recalled echo-based MR elastography at 3 Tesla with and without gadoxetic acid administration

Affiliations
Comparative Study

Comparison of spin-echo echoplanar imaging and gradient recalled echo-based MR elastography at 3 Tesla with and without gadoxetic acid administration

Yong Seek Kim et al. Eur Radiol. 2017 Oct.

Abstract

Objective: To compare spin-echo echoplanar imaging (SE-EPI) and gradient recalled echo (GRE) MR elastography (MRE) at 3 T with and without gadoxetic acid administration.

Methods: We included 84 patients who underwent MRE before and after gadoxetic acid administration, each time using SE-EPI and GRE sequences. Diagnostic performance for predicting clinical liver cirrhosis and high-risk oesophageal varices was assessed using the area under the receiver-operating characteristic curve (AUC). The relationships between T2* and success of MRE, and correlations of liver stiffness (LS) values between the two sequences or before and after gadoxetic acid administration, were investigated.

Results: SE-EPI-MRE resulted in a significantly lower failure rate than GRE-MRE (1.19% vs. 10.71%, P = 0.018). Increased T2* was related to higher probability of successful LS measurement (odds ratio, 1.426; P = 0.004). The AUC of SE-EPI-MRE was comparable to that of GRE-MRE for the detection of clinical liver cirrhosis (0.938 vs. 0.948, P = 0.235) and high-risk oesophageal varices (0.839 vs. 0.752, P = 0.354). LS values were not significantly different before and after gadoxetic acid administration.

Conclusion: SE-EPI-MRE can substitute for GRE-MRE for the detection of clinical liver cirrhosis and high-risk oesophageal varices. SE-EPI-MRE is particularly useful in patients with iron deposition, with lower failure rates than GRE-MRE.

Key points: • LS values are comparable between SE-EPI-MRE and GRE-MRE. • Administration of gadoxetic acid does not influence LS measurement. • The failure rate of SE-EPI-MRE is significantly lower than that of GRE-MRE.

Keywords: Gadoxetic acid; Gradient recalled echo; Liver fibrosis; Magnetic resonance elastography; Spin echo.

PubMed Disclaimer

References

    1. Magn Reson Imaging. 2012 Jan;30(1):128-32 - PubMed
    1. Radiology. 2016 Jan;278(1):114-24 - PubMed
    1. World J Gastroenterol. 2014 Dec 14;20(46):17558-67 - PubMed
    1. Eur Radiol. 2008 Nov;18(11):2535-41 - PubMed
    1. J Magn Reson Imaging. 2016 Sep;44(3):715-22 - PubMed

Publication types

LinkOut - more resources