The Accelograph: a new neuromuscular transmission monitor
- PMID: 2830761
- DOI: 10.1111/j.1399-6576.1988.tb02687.x
The Accelograph: a new neuromuscular transmission monitor
Abstract
A new neuromuscular transmission monitor based on measurement of acceleration is presented. The monitor consists of an acceleration transducer, i.e. a piezo-electric ceramic wafer with an electrode on each side, and a stimulation and computing unit. The transducer can be fastened to the thumb, and when the finger is moved in response to nerve stimulation a voltage difference develops between the two electrodes. The voltage is then measured and registered in the computing unit. The acceleration transducer does not require preload, but merely that the muscle on which the measurement is being made can move freely. The accelograph is therefore easy to set up and operate. In daily clinical practice it has proved possible, using this new apparatus, to achieve reliable evaluation of the degree of non-depolarizing neuromuscular blockade in less than 2 min.
Similar articles
-
Measurement of acceleration: a new method of monitoring neuromuscular function.Acta Anaesthesiol Scand. 1988 Jan;32(1):45-8. doi: 10.1111/j.1399-6576.1988.tb02686.x. Acta Anaesthesiol Scand. 1988. PMID: 2830760
-
[A new neuromuscular transmission monitor (TOF Guard): the rationale behind the method and its clinical usefulness].Masui. 1994 Jan;43(1):134-9. Masui. 1994. PMID: 8309049 Japanese.
-
[The design and clinical application of an apparatus for monitoring neuromuscular transmission function during operation under general anaesthesia].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 1998 Jun;15(2):123-7. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 1998. PMID: 12548897 Chinese.
-
Monitoring of neuromuscular transmission by electromyography during anaesthesia. A comparison with mechanomyography in cat and man.Dan Med Bull. 1996 Sep;43(4):301-16. Dan Med Bull. 1996. PMID: 8884132 Review.
-
[Value of monitoring muscle relaxation].Schweiz Med Wochenschr. 1996 Sep 28;126(39):1649-53. Schweiz Med Wochenschr. 1996. PMID: 8927968 Review. German.
Cited by
-
Monitoring neuromuscular function in the intensive care unit.Intensive Care Med. 1993;19 Suppl 2:S74-9. doi: 10.1007/BF01708805. Intensive Care Med. 1993. PMID: 8106682 Review. No abstract available.
-
Efficacy and safety of sugammadex versus neostigmine in reversing neuromuscular blockade in adults.Cochrane Database Syst Rev. 2017 Aug 14;8(8):CD012763. doi: 10.1002/14651858.CD012763. Cochrane Database Syst Rev. 2017. PMID: 28806470 Free PMC article.
-
Clinical assessment of a new neuromuscular transmission monitoring system (ACCELOGRAPH--a comparison with the conventional method.J Anesth. 1989 Mar 1;3(1):90-3. doi: 10.1007/s0054090030090. J Anesth. 1989. PMID: 15236063 No abstract available.
-
Effect of sevoflurane anesthesia on neuromuscular blockade produced by rocuronium infusion in dogs.J Vet Med Sci. 2019 Mar 20;81(3):425-433. doi: 10.1292/jvms.18-0479. Epub 2019 Jan 30. J Vet Med Sci. 2019. PMID: 30700651 Free PMC article.
-
Neuromuscular pharmacology update.J Anesth. 1992 Jul;6(3):332-40. doi: 10.1007/s0054020060332. J Anesth. 1992. PMID: 15278545 No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources