Creatinine Assay Attainment of Analytical Performance Goals Following Implementation of IDMS Standardization: Further Improvements Required
- PMID: 28321322
- PMCID: PMC5347424
- DOI: 10.1177/2054358117693353
Creatinine Assay Attainment of Analytical Performance Goals Following Implementation of IDMS Standardization: Further Improvements Required
Abstract
Background: The international initiative to standardize creatinine (Cr) assays by tracing reference materials to Isotope Dilution Mass Spectrometry (IDMS) assigned values was implemented to reduce interlaboratory variability and improve assay accuracy.
Objective: The aims of this study were to examine whether IDMS standardization has improved Cr assay accuracy (bias), interlaboratory variability (precision), total error (TE), and attainment of recommended analytical performance goals.
Methods: External Quality Assessment (EQA) data (n = 66 challenge vials) from Ontario, Canada, were analyzed. The bias, precision, TE, and the number of EQA challenge vials meeting performance goals were determined by assay manufacturer before (n = 32) and after (n = 34) IDMS implementation.
Results: The challenge vials with the worst bias and precision were spiked with known common interfering substances (glucose and bilirubin). IDMS standardization improved assay bias (10.4%-1.6%, P < .001), but precision remained unchanged (5.0%-4.7%, P = .5) with performance goals not consistently being met. Precision and TE goals based on biologic variation were attained by only 29% to 69% and 32% to 62% of challenge vials.
Conclusions: While IDMS standardization has improved Cr assay accuracy and thus reduced TE, significant interlaboratory variability remains. Contemporary Cr assays do not currently meet the standards required to allow for accurate and consistent estimated glomerular filtration rate assessment and chronic kidney disease diagnosis across laboratories. Further improvements in Cr assay performance are needed.
Mise en contexte: L’initiative internationale visant à standardiser les épreuves de dosage de la créatinine par le traçage des substances de référence aux valeurs qui leur sont assignées par spectrométrie de masse à dilution isotopique (IDMS) a été mise en œuvre dans le but de réduire la variabilité inter-laboratoires et d’améliorer la précision de l’essai lui-même.
Objectifs de l’étude: Cette étude visait à déterminer si la standardisation par IDMS parvenait à améliorer l’exactitude de l’épreuve de dosage de la créatinine (biais), à réduire la variabilité inter-laboratoires (précision) et si elle contribuait à l’atteinte des objectifs recommandés en matière de performance analytique.
Méthodologie: On a analysé les données de l’assurance externe de la qualité (AEQ) provenant de l’Ontario, au Canada. Le biais, l’exactitude, l’erreur totale ainsi que le nombre de fioles d’épreuves provenant du AEQ atteignant les objectifs d’erreurs ont été déterminés par le fabricant de l’épreuve avant et après la mise en œuvre de l’IDMS.
Résultats: Les essais enzymatiques ont obtenu des résultats plus précis et des valeurs d’erreur totale inférieures. Les fioles d’épreuves qui présentaient la plus faible précision et le plus fort biais étaient celles à qui l’on avait ajouté des substances connues pour créer de l’interférence (glucose et bilirubine). La standardisation par IDMS a permis d’améliorer le biais lié à l’épreuve (de 10,4% à 1,6%, P < 0,001), mais n’a pas affecté la précision qui est demeurée faible (5,0 à 4,7%, P = 0,5). De plus, les objectifs de performance n’ont pas toujours été atteints: les objectifs de précision et d’erreur totale établis en tenant compte des variations biologiques ont été atteints par seulement 29 à 69% et 32 à 62% des fioles d’épreuves.
Conclusions: Bien que la standardisation par IDMS ait réduit les valeurs d’erreur totale obtenues lors des épreuves de dosage de la créatinine et ait amélioré la précision de l’essai, une variabilité inter-laboratoires significative subsiste dans les résultats. À ce jour, les épreuves de dosage de la créatinine ne rencontrent toujours pas les standards exigés pour permettre une mesure précise et homogène du débit de filtration glomérulaire estimé et le diagnostic de l’insuffisance rénale chronique dans tous les laboratoires. Par conséquent, des améliorations supplémentaires en ce qui concerne la performance des épreuves de dosage de la créatinine s’avèrent nécessaires.
Keywords: External Quality Assessment; IDMS traceable; chronic kidney disease; creatinine.
Conflict of interest statement
Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.
Figures

Similar articles
-
Interlaboratory Variability in Plasma Creatinine Measurement and the Relation with Estimated Glomerular Filtration Rate and Chronic Kidney Disease Diagnosis.Clin J Am Soc Nephrol. 2017 Jan 6;12(1):29-37. doi: 10.2215/CJN.05400516. Epub 2016 Nov 8. Clin J Am Soc Nephrol. 2017. PMID: 27827312 Free PMC article.
-
Results of the recalibration of creatinine measurement with the modular Beckman Coulter Jaffe creatinine method.Clin Chem Lab Med. 2011 Dec;49(12):1987-99. doi: 10.1515/CCLM.2011.674. Epub 2011 Jul 29. Clin Chem Lab Med. 2011. PMID: 21801031
-
Modification of the Modification of Diet in Renal Disease (MDRD) Study equation for Japan.Am J Kidney Dis. 2007 Dec;50(6):927-37. doi: 10.1053/j.ajkd.2007.09.004. Am J Kidney Dis. 2007. PMID: 18037093
-
Calibration and precision of serum creatinine and plasma cystatin C measurement: impact on the estimation of glomerular filtration rate.J Nephrol. 2014 Oct;27(5):467-75. doi: 10.1007/s40620-014-0087-7. Epub 2014 Apr 8. J Nephrol. 2014. PMID: 24711159 Review.
-
Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program.Clin Chem. 2006 Jan;52(1):5-18. doi: 10.1373/clinchem.2005.0525144. Epub 2005 Dec 6. Clin Chem. 2006. PMID: 16332993 Review.
Cited by
-
Individual patient variability with the application of the kidney failure risk equation in advanced chronic kidney disease.PLoS One. 2018 Jun 12;13(6):e0198456. doi: 10.1371/journal.pone.0198456. eCollection 2018. PLoS One. 2018. PMID: 29894480 Free PMC article.
-
Creatinine standardization: a key consideration in evaluating whole blood creatinine monitoring systems for CKD screening.Anal Bioanal Chem. 2022 Apr;414(10):3279-3289. doi: 10.1007/s00216-022-03942-7. Epub 2022 Mar 9. Anal Bioanal Chem. 2022. PMID: 35260937 Free PMC article.
-
Conversion methods for modified Jaffe reaction assays of serum creatinine.Porto Biomed J. 2020 Jun 4;5(3):e72. doi: 10.1097/j.pbj.0000000000000072. eCollection 2020 May-Jun. Porto Biomed J. 2020. PMID: 33299949 Free PMC article.
References
-
- Kidney Disease: Improving Global Outcomes (KDIGO) Work Group. KDIGO 2012 clinical practice guideline for the evaluation and management of chronic kidney disease. Kidney Int. 2012;3:1-150. - PubMed
-
- Levey AS. Measurement of renal function in chronic renal disease. Kidney Int. 1990;38:167-184. - PubMed
-
- Myers GL, Miller WG, Coresh J, et al. Recommendations for improving serum creatinine measurement: a report from the Laboratory Working Group of the National Kidney Disease Education Program. Clin Chem. 2006;52(1):5-18. - PubMed
-
- Miller WG, Myers GL, Ashwood ER, et al. Creatinine measurement: state of the art in accuracy and interlaboratory harmonization. Arch Pathol Lab Med. 2005;129(3):297-304. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous