On the reproducibility of expert-operated and robotic ultrasound acquisitions
- PMID: 28321804
- DOI: 10.1007/s11548-017-1561-1
On the reproducibility of expert-operated and robotic ultrasound acquisitions
Erratum in
-
Correction to: On the reproducibility of expert-operated and robotic ultrasound acquisitions.Int J Comput Assist Radiol Surg. 2019 Oct;14(10):1821. doi: 10.1007/s11548-019-02055-4. Int J Comput Assist Radiol Surg. 2019. PMID: 31432359
Abstract
Purpose: We present the evaluation of the reproducibility of measurements performed using robotic ultrasound imaging in comparison with expert-operated sonography. Robotic imaging for interventional procedures may be a valuable contribution, but requires reproducibility for its acceptance in clinical routine. We study this by comparing repeated measurements based on robotic and expert-operated ultrasound imaging.
Methods: Robotic ultrasound acquisition is performed in three steps under user guidance: First, the patient is observed using a 3D camera on the robot end effector, and the user selects the region of interest. This allows for automatic planning of the robot trajectory. Next, the robot executes a sweeping motion following the planned trajectory, during which the ultrasound images and tracking data are recorded. As the robot is compliant, deviations from the path are possible, for instance due to patient motion. Finally, the ultrasound slices are compounded to create a volume. Repeated acquisitions can be performed automatically by comparing the previous and current patient surface.
Results: After repeated image acquisitions, the measurements based on acquisitions performed by the robotic system and expert are compared. Within our case series, the expert measured the anterior-posterior, longitudinal, transversal lengths of both of the left and right thyroid lobes on each of the 4 healthy volunteers 3 times, providing 72 measurements. Subsequently, the same procedure was performed using the robotic system resulting in a cumulative total of 144 clinically relevant measurements. Our results clearly indicated that robotic ultrasound enables more repeatable measurements.
Conclusions: A robotic ultrasound platform leads to more reproducible data, which is of crucial importance for planning and executing interventions.
Keywords: Autonomous acquisition; Reproducibility; Robotic ultrasound; Thyroid; Ultrasound acquisition.
Similar articles
-
Robotic ultrasound imaging: State-of-the-art and future perspectives.Med Image Anal. 2023 Oct;89:102878. doi: 10.1016/j.media.2023.102878. Epub 2023 Jul 18. Med Image Anal. 2023. PMID: 37541100 Review.
-
Towards MRI-Based Autonomous Robotic US Acquisitions: A First Feasibility Study.IEEE Trans Med Imaging. 2017 Feb;36(2):538-548. doi: 10.1109/TMI.2016.2620723. Epub 2016 Oct 24. IEEE Trans Med Imaging. 2017. PMID: 27831861
-
Validation of a novel robot-assisted 3DUS system for real-time planning and guidance of breast interstitial HDR brachytherapy.Med Phys. 2015 Dec;42(12):6830-9. doi: 10.1118/1.4934832. Med Phys. 2015. PMID: 26632040
-
Acoustic window planning for ultrasound acquisition.Int J Comput Assist Radiol Surg. 2017 Jun;12(6):993-1001. doi: 10.1007/s11548-017-1551-3. Epub 2017 Mar 11. Int J Comput Assist Radiol Surg. 2017. PMID: 28285339 Free PMC article.
-
Robot-assisted ultrasound imaging: overview and development of a parallel telerobotic system.Minim Invasive Ther Allied Technol. 2015 Feb;24(1):54-62. doi: 10.3109/13645706.2014.992908. Epub 2014 Dec 25. Minim Invasive Ther Allied Technol. 2015. PMID: 25540071 Review.
Cited by
-
Autonomous robotic ultrasound scanning system: a key to enhancing image analysis reproducibility and observer consistency in ultrasound imaging.Front Robot AI. 2025 Feb 5;12:1527686. doi: 10.3389/frobt.2025.1527686. eCollection 2025. Front Robot AI. 2025. PMID: 39975565 Free PMC article.
-
Review of robotic systems for thoracoabdominal puncture interventional surgery.APL Bioeng. 2024 Apr 1;8(2):021501. doi: 10.1063/5.0180494. eCollection 2024 Jun. APL Bioeng. 2024. PMID: 38572313 Free PMC article. Review.
-
Robotic Assisted Ultrasound-Guided Endovascular Stent Implantation in a Vascular Model.Int J Med Robot. 2024 Dec;20(6):e70005. doi: 10.1002/rcs.70005. Int J Med Robot. 2024. PMID: 39501897 Free PMC article.
-
Robotic Ultrasound Scanning End-Effector with Adjustable Constant Contact Force.Cyborg Bionic Syst. 2025 May 2;6:0251. doi: 10.34133/cbsystems.0251. eCollection 2025. Cyborg Bionic Syst. 2025. PMID: 40321899 Free PMC article.
-
Towards autonomous robotic THz-based in vivo skin sensing: the PicoBot.Sci Rep. 2025 Feb 7;15(1):4568. doi: 10.1038/s41598-025-88718-6. Sci Rep. 2025. PMID: 39915605 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials