Pushing the limits: establishing an ultra-low framerate and antiscatter grid-less radiation protocol for left atrial ablations
Abstract
Aims: Despite the use of 3D mapping systems and new developments of non-fluoroscopic options, most centres still rely at least in part on fluoroscopy for catheter visualization during catheter ablations. The purpose of this study was to assess the feasibility of using an ultra-low frame rate and antiscatter grid-less radiation protocol during complex left atrial ablations to minimize radiation exposure for the patient and staff.
Methods and results: A total of 150 consecutive patients undergoing left atrial ablations in our hospital were included in the analysis. The procedures were performed between January 2015 and November 2016. Of the included patients 75 (50%) underwent ablation before and 75 (50%) after the ultra-low frame rate (reduced from 4 to 2 FPS) and antiscatter grid-less radiation protocol was established. Procedures performed after the dose reduction protocol was established showed a 64% reduction of the dose area product (630.28 ± 550.96 vs. 226.44 ± 277.44 µGym2, P < 0.001), while fluoroscopy duration (14.22 ± 4.47 vs. 13.62 ± 7.11 min, P = 0.066) and procedural duration (1:48 ± 0:28 vs. 1:53 ± 0:34 min, P = 0.525) were not prolonged. Acute procedural success was achieved in all procedures. Two complications occurred before and one complication after the protocol was established. During four procedures, operators decided to re-introduce the antiscatter grid. This was due to impaired visibility in morbidly obese patients (n = 2) or technically difficult transseptal puncture (n = 2).
Conclusion: The use of an ultra low framerate and antiscatter grid-less radiation protocol effectively reduced radiation dose for complex left atrial ablation procedures and lead to very low average patient doses. Reduced image quality did not impair procedural and fluoroscopy duration or acute procedural success.
Similar articles
-
Safe procedures despite ultra low radiation doses during catheter ablations of atrial and ventricular arrhythmias-A multicenter experience.Pacing Clin Electrophysiol. 2021 May;44(5):807-813. doi: 10.1111/pace.14205. Epub 2021 Mar 15. Pacing Clin Electrophysiol. 2021. PMID: 33665850
-
Safety profile of near-zero fluoroscopy atrial fibrillation ablation with non-fluoroscopic catheter visualization: experience from 1000 consecutive procedures.Europace. 2018 Dec 1;20(12):1952-1958. doi: 10.1093/europace/eux378. Europace. 2018. PMID: 29346552
-
Use of a contact force-sensing ablation catheter with advanced catheter location significantly reduces fluoroscopy time and radiation dose in catheter ablation of atrial fibrillation.Europace. 2016 Feb;18(2):211-8. doi: 10.1093/europace/euv186. Epub 2015 Aug 27. Europace. 2016. PMID: 26318548
-
Meta-analysis of pulmonary vein isolation ablation for atrial fibrillation conventional vs low- and zero-fluoroscopy approaches.J Cardiovasc Electrophysiol. 2020 Jun;31(6):1403-1412. doi: 10.1111/jce.14450. Epub 2020 Mar 30. J Cardiovasc Electrophysiol. 2020. PMID: 32202003
-
Three-dimensional localization versus fluoroscopically only guided ablations: a meta-analysis.Scand Cardiovasc J. 2013 Aug;47(4):200-9. doi: 10.3109/14017431.2013.797099. Epub 2013 May 27. Scand Cardiovasc J. 2013. PMID: 23647246 Review.
Cited by
-
Safety and Effectiveness of a Novel Fluoroless Transseptal Puncture Technique for Lead-free Catheter Ablation: A Case Series.J Innov Card Rhythm Manag. 2020 Apr 15;11(4):4079-4085. doi: 10.19102/icrm.2020.110405. eCollection 2020 Apr. J Innov Card Rhythm Manag. 2020. PMID: 32368383 Free PMC article.
-
[Radiation reduction in interventional electrophysiology : Results from operators with different levels of experience].Herzschrittmacherther Elektrophysiol. 2018 Dec;29(4):406-410. doi: 10.1007/s00399-018-0589-4. Herzschrittmacherther Elektrophysiol. 2018. PMID: 30155562 German.
-
Fluoroless left atrial access for radiofrequency and cryoballoon ablations using a novel radiofrequency transseptal wire.J Interv Card Electrophysiol. 2022 Jun;64(1):183-190. doi: 10.1007/s10840-022-01157-5. Epub 2022 Feb 22. J Interv Card Electrophysiol. 2022. PMID: 35194727 Free PMC article.
-
Minimizing Radiation in the Modern Electrophysiology Laboratory.J Innov Card Rhythm Manag. 2018 Aug 15;9(8):3265-3270. doi: 10.19102/icrm.2018.090805. eCollection 2018 Aug. J Innov Card Rhythm Manag. 2018. PMID: 32494501 Free PMC article. Review.
-
Results from a real-time dosimetry study during left atrial ablations performed with ultra-low dose radiation settings.Herzschrittmacherther Elektrophysiol. 2021 Jun;32(2):244-249. doi: 10.1007/s00399-021-00762-7. Epub 2021 May 11. Herzschrittmacherther Elektrophysiol. 2021. PMID: 33977306 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous