Aorto-Right Atrial Fistula After Sutureless Valve Implantation
- PMID: 28347580
- DOI: 10.1016/j.cjca.2017.01.009
Aorto-Right Atrial Fistula After Sutureless Valve Implantation
Abstract
Aortic valve replacement with sutureless valves has many potential applications including in redo surgery, minimally invasive scenarios, and heavily calcified aortic roots. Herein we report a case of the development of an aorto-right atrial fistula after replacement of a Medtronic Freestyle stentless subcoronary bioprosthesis (Medtronic Inc, Minneapolis, MN) with a Perceval sutureless valve (LivaNova PLC, London, UK). This eventually necessitated repair with repeat surgery and aortic valve replacement with a stented valve. For patients with failing stentless bioprostheses returning for reintervention, we suggest removal of only Freestyle leaflet tissue without supra-annular debridement to avoid weakening of the native root tissue.
Copyright © 2017 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.