Predictors of Recurrent Atrial Fibrillation Using Mode Switch Quantification
- PMID: 28352435
- PMCID: PMC5358197
- DOI: 10.4021/cr292w
Predictors of Recurrent Atrial Fibrillation Using Mode Switch Quantification
Abstract
Background: The efficacy of mode switching to predict atrial fibrillation has been established in the literature. There have been few studies investigating the incidence and clinical implication of mode switch episodes quantified from implantable cardioverter defibrillator and pacemaker interrogation. We sought to investigate the incidence of mode switch recurrence in patients with implantable cardioverter defibrillators and permanent pacemakers.
Methods: Mode switch was defined as any occurrence documented during device interrogation after the date of implantation. Clinical predictors (age, gender, hypertension, diabetes, syncope, atrial fibrillation (AF)), and medications were analyzed to determine association with single and recurrent mode switch occurrences.
Results: There were 21 patients experiencing a mode switch event, identified from a group of 54 patients (42 males; mean age 70 ± 12 years; mean follow-up 29.1 ± 22 months (3.4 - 81.4 months)). All but two patients were receiving medical therapy including beta blockers, statins, ace-inhibitors, and anti-arrhythmics. There were 21 subjects who experienced at least one mode switch during their follow-up and 33 subjects who never experienced a mode switch during their follow-up time. The median time to first mode switch from device implantation was 39.3 months. Risk factors individually associated with any mode switch episode included: diabetes (DM) (P < 0.04) and use of digitalis (P = 0.02). Subjects who had a history of DM were 5 times more likely to have at least one mode switch occurrence. There was a significantly higher rate of mode switch among patients who were diabetic than patients who were not (3.7 per follow-up month ± 5.3 vs. 0.98 per follow-up month ± 2.02; P = 0.02). There was a significantly higher rate of mode switch among patients who were on digitalis than those who were not (3.1 per follow-up month ± 4.3 vs. 0.73 per follow-up month ± 1.9; P = 0.02).
Conclusion: The main factors associated with any mode switch are having a history of diabetes and digitalis use. Those patients who are diabetics and those on digitalis may warrant closer observation and management for the development of atrial fibrillation.
Keywords: Atrial fibrillation; Cardioverter defibrillator; Mode switch; Permanent pacemaker.
Similar articles
-
Clinical Implications of Brief Device-Detected Atrial Tachyarrhythmias in a Cardiac Rhythm Management Device Population: Results from the Registry of Atrial Tachycardia and Atrial Fibrillation Episodes.Circulation. 2016 Oct 18;134(16):1130-1140. doi: 10.1161/CIRCULATIONAHA.115.020252. Circulation. 2016. PMID: 27754946
-
Atrial Rate-Responsive Pacing and Incidence of Sustained Atrial Arrhythmias in Patients with Implantable Cardioverter Defibrillators.Pacing Clin Electrophysiol. 2016 Jun;39(6):548-56. doi: 10.1111/pace.12856. Epub 2016 Apr 22. Pacing Clin Electrophysiol. 2016. PMID: 27001125
-
Reduction of atrial fibrillation in remotely monitored pacemaker patients: results from a Chinese multicentre registry.Chin Med J (Engl). 2013 Nov;126(22):4216-21. Chin Med J (Engl). 2013. PMID: 24238500
-
Prediction of atrial fibrillation in patients with an implantable cardioverter-defibrillator and heart failure.Eur J Heart Fail. 2010 Oct;12(10):1101-10. doi: 10.1093/eurjhf/hfq126. Eur J Heart Fail. 2010. PMID: 20861134
-
[Pacemaker therapy in patients with atrial fibrillation].Herz. 1998 Jun;23(4):260-8. doi: 10.1007/BF03044321. Herz. 1998. PMID: 9690113 Review. German.
References
-
- Bernstein AD, Daubert JC, Fletcher RD, Hayes DL, Luderitz B, Reynolds DW, Schoenfeld MH. et al. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology/British Pacing and Electrophysiology Group. Pacing Clin Electrophysiol. 2002;25(2):260–264. doi: 10.1046/j.1460-9592.2002.00260.x. - DOI - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous