Comparison of four electronic root canal length measurement devices
Abstract
Objective: The objective of this study is to evaluate the performance of four electronic root canal length measurement devices (ERCLMDs) [Root ZX, Raypex 6, ProPex II, and VDW Gold with ERCLMD] in wet and dry root canals with different major foramen diameters, in association with three file positions within the final 1 mm of canals.
Materials and methods: Fifty roots were divided into five groups that were instrumented apically to the terminus diameter using K-files 15, 20, 30, 40, and 60. ERCLMD measurements were made when the file tip was at major foramen, 0.5 mm and 1.0 mm short of the foramen, when the apex touched to the surface of saline or was immersed 3 mm into saline, and when the canal was dry or was irrigated with saline. The differences between the electronic lengths and the actual lengths were calculated. The data were analyzed using the three-, two-, and one-way ANOVA and the LSD test (p < .05).
Results: Overall, the four ERCLMDs demonstrated 12.8% unstable and beyond the foramen measurements (11.3%). About 85% of the beyond the foramen measurements had foramen diameters 0.40 and 0.60. ERCLMD comparisons related to the file positions in the foramen diameters showed significant differences between ERCLMDs in each foramen diameter (p < .05).
Conclusions: All ERCLMDs provided highly accurate measurements within the final 1 mm of the foramen. Wet or dry canals and apex conditions did not adversely affect the accuracy of the ERCLMDs' readings. ProPex II located the file positions in the teeth with different foramen diameters more accurately than the other ERCLMDs.
Keywords: Electronic root canal length measurement device; apical foramen; apical foramen diameter; working length.
Similar articles
-
In vivo accuracy of three electronic root canal length measurement devices: Dentaport ZX, Raypex 5 and ProPex II.Int Endod J. 2012 Jun;45(6):552-6. doi: 10.1111/j.1365-2591.2011.02010.x. Epub 2012 Jan 19. Int Endod J. 2012. PMID: 22257341 Clinical Trial.
-
Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.Clin Oral Investig. 2018 Dec;22(9):3107-3112. doi: 10.1007/s00784-018-2404-3. Epub 2018 Mar 7. Clin Oral Investig. 2018. PMID: 29516188
-
Influence of instrument size and varying electrical resistance of root canal instruments on accuracy of three electronic root canal length measurement devices.Int Endod J. 2017 May;50(5):506-511. doi: 10.1111/iej.12649. Epub 2016 May 5. Int Endod J. 2017. PMID: 27079789
-
The fundamental operating principles of electronic root canal length measurement devices.Int Endod J. 2006 Aug;39(8):595-609. doi: 10.1111/j.1365-2591.2006.01131.x. Int Endod J. 2006. PMID: 16872454 Review.
-
Electronic canal length measurement.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995 Feb;79(2):226-31. doi: 10.1016/s1079-2104(05)80288-4. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995. PMID: 7614187 Review.
Cited by
-
The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.Clin Oral Investig. 2021 Sep;25(9):5239-5245. doi: 10.1007/s00784-021-03831-w. Epub 2021 Feb 13. Clin Oral Investig. 2021. PMID: 33580848
-
How Do Different Image Modules Impact the Accuracy of Working Length Measurements in Digital Periapical Radiography? An In Vitro Study.Diagnostics (Basel). 2025 Jan 28;15(3):305. doi: 10.3390/diagnostics15030305. Diagnostics (Basel). 2025. PMID: 39941235 Free PMC article.
-
Factors influencing the accuracy of electronic apex locators: A scoping review.Dent Res J (Isfahan). 2025 Jun 27;22:26. doi: 10.4103/drj.drj_634_23. eCollection 2025. Dent Res J (Isfahan). 2025. PMID: 40655901 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous