Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2017 May;18(3):73-82.
doi: 10.1002/acm2.12074. Epub 2017 Mar 29.

Validation of a GPU-Based 3D dose calculator for modulated beams

Affiliations

Validation of a GPU-Based 3D dose calculator for modulated beams

Saeed Ahmed et al. J Appl Clin Med Phys. 2017 May.

Abstract

A superposition/convolution GPU-accelerated dose computation algorithm (the Calculator) has been recently incorporated into commercial software. The algorithm requires validation prior to clinical use. Three photon energies were examined: conventional 6 MV and 15 MV, and 10 MV flattening filter free (10 MVFFF). For a set of IMRT and VMAT plans based on four of the five AAPM Practice Guideline 5a downloadable datasets, ion chamber (IC) measurements were performed on the water-equivalent phantoms. The average difference between the Calculator and IC was -0.3 ± 0.8% (1SD). The same plans were projected on a phantom containing a biplanar diode array. We used the forthcoming criteria for routine gamma analysis, 3% dose-error (global (G) normalization, 2 mm distance to agreement, and 10% low dose cutoff). The γ (3%G/2 mm) average passing rate was 98.9 ± 2.1%. Measurement-guided three-dimensional dose reconstruction on the patient CT dataset (excluding the Lung) resulted in a similar average agreement rate with the Calculator: 98.2 ± 2.0%. The mean γ (3%G/2 mm) passing rate comparing the Calculator to the TPS (again excluding the Lung) was 99.0 ± 1.0%. Because of the significant inhomogeneity, the Lung case was investigated separately. The calculator has an alternate heterogeneity correction mode that can change the results in the thorax for higher-energy beams (15 MV). As this correction is nonphysical and was optimized for simple slab geometries, its application leads to mixed results when compared to the TPS and independent Monte Carlo calculations, depending on the CT dataset and the plan. The Calculator vs. TPS 15 MV Guideline 5a IMRT and VMAT plans demonstrate 96.3% and 93.4% γ (3%G/2 mm) passing rates respectively. For the lower energies, which should be predominantly used in the thoracic region, the passing rates for the same plans and criteria range from 98.6 to 100%. Overall, the Calculator accuracy is sufficient for the intended use.

Keywords: GPU-accelerated calculations; convolution/superposition; dose calculations; segmented beams.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Gamma analysis (2%L/2 mm) error maps (inserts) between Pinnacle and SNC calculator, and normalized cross‐plane dose profiles at 20 cm depth for 10MVFFF (a) and 15 MV (b) 20 × 20 cm2 fields. Ion chamber (IC) profiles are also included for comparison. Red and blue pixels are where the Calculator dose is above and below Pinnacle respectively.
Figure 2
Figure 2
Central axis PDDs on the lung slab phantom for the 2 × 2 cm2 field: 15MV (a), 10 MVFFF (b), and 6MV (c).
Figure 3
Figure 3
Frequency distribution of the γ‐analysis passing rates comparing the Calculator to the Delta4 for all Guideline 5a test cases and energies. N = 24.
Figure 4
Figure 4
Frequency distribution of the γ‐analysis passing rates comparing the Calculator to ACPDP measurement‐guided dose reconstruction on the patient datasets for Guideline 5a test cases excluding the Lung (all energies). N = 18.
Figure 5
Figure 5
Graphical representation of the γ(2%G/2 mm) comparison for the 6X Anal IMRT plan between the Calculator and ACPDP (left column) and Pinnacle vs. ACPDP (right column). The highlighted pixels where the gamma analysis fails are overlaid on the dose map. The targets are also shown. The transverse cuts (a), (b) are taken superiorly to demonstrate the areas of failure between the irradiated nodal chains. The coronal cuts (c), (d) are 2 cm posterior to the midline, where both the primary and secondary targets are prominently present.
Figure 6
Figure 6
Frequency distribution of the γ‐analysis passing rates comparing the Calculator to Pinnacle on the patient datasets for Guideline 5a test cases, excluding the Lung (all energies). N = 18.

References

    1. ACR ASTRO Practice guideline for intensity modulated radiation therapy (IMRT). 2011. http://www.acr.org/~/media/ACR/Documents/PGTS/guidelines/IMRT.pdf accessed 01/25/2017.
    1. Stern RL, Heaton R, Fraser MW, et al. Verification of monitor unit calculations for non‐IMRT clinical radiotherapy: report of AAPM Task Group 114. Med Phys. 2011;38:504–530. - PubMed
    1. Low DA, Moran JM, Dempsey JF, et al. Dosimetry tools and techniques for IMRT. Med Phys. 2011;38:1313. - PubMed
    1. van Elmpt W, McDermott L, Nijsten S, et al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother Oncol. 2008;88:289–309. - PubMed
    1. van Elmpt W, Nijsten S, Mijnheer B, et al. The next step in patient‐specific QA: 3D dose verification of conformal and intensity‐modulated RT based on EPID dosimetry and Monte Carlo dose calculations. Radiother Oncol. 2008;86:86–92. - PubMed

Publication types

LinkOut - more resources