Viability of Hand and Wrist Photogoniometry
- PMID: 28391753
- PMCID: PMC5987967
- DOI: 10.1177/1558944717702471
Viability of Hand and Wrist Photogoniometry
Abstract
Background: No goniometric technique is both maximally convenient and completely accurate, although photogoniometry (ie, picture taking to facilitate digital angle measurement) shows promise in this regard. Our purpose was to test the feasibility and reliability of a photogoniometric protocol designed to measure wrist and digit range of motion in general.
Methods: Two independent observers examined a sample of joints in both normal and abnormal hands according to a photogoniometric protocol. Interrater and intrarater correlation were calculated, and these measurements were compared with measurements made by a third independent examiner with a manual goniometer.
Results: The photo-based measurements were reliable within and between observers; however, only a minority of these measurements were in agreement with manually collected values.
Conclusions: At present, photogoniometry is not an acceptable alternative to manual goniometry for determining wrist and digit range of motion in general. Joint-specific photogoniometry should be the subject of future study, as should relevant imaging and software technology.
Keywords: digit range of motion; digital angle measurement; manual goniometry; photogoniometry; wrist range of motion.
Conflict of interest statement
References
-
- Acheson RM, Egnatz D, Roess M. A photographic device for recording progress in arthritis of the hands. Arthritis Rheum. 1966;9:72-73. - PubMed
-
- Boone DC, Azen SP, Lin CM, et al. Reliability of goniometric measurements. Phys Ther. 1978;58:1355-1360. - PubMed
-
- Carpinella I, Mazzoleni P, Rabuffetti M, et al. Experimental protocol for the kinematic analysis of the hand: definition and repeatability. Gait Posture. 2006;23:445-454. - PubMed
-
- Carter TI, Pansy B, Wolff AL, et al. Accuracy and reliability of three different techniques for manual goniometry for wrist motion: a cadaveric study. J Hand Surg Am. 2009;34:1422-1428. - PubMed
-
- Charlton PC, Mentiplay BF, Pua YH, et al. Reliability and concurrent validity of a Smartphone, bubble inclinometer and motion analysis system for measurement of hip joint range of motion. J Sci Med Sport. 2015;18:262-267. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources