Abstract
Objectives: To investigate whether (1) the type of fixation at the shaft (hybrid vs. locking), (2) the position of the plate (offset vs. contact) and (3) the implant material has a significant effect on (a) construct stiffness and (b) fatigue life in a distal femur extraarticular comminuted fracture model using the same design of distal femur periarticular locking plate.
Methods: An extraarticular severely comminuted distal femoral fracture pattern (OTA/AO 33-A3) was simulated using artificial bone substitutes. Ten-hole distal lateral femur locking plates were used for fixation per the recommended surgical technique. At the distal metaphyseal fragment, all possible locking screws were placed. For the proximal diaphyseal fragment, different types of screws were used to create 4 different fixation constructs: (1) stainless steel hybrid (SSH), (2) stainless steel locked (SSL), (3) titanium locked (TiL), and (4) stainless steel locked with 5-mm offset at the diaphysis (SSLO). Six specimens of each construct configuration were tested. First, each specimen was nondestructively loaded axially to determine the stiffness. Then, each specimen was cyclically loaded with increasing load levels until failure.
Results: Construct Stiffness: The fixation construct with a stainless steel plate and hybrid fixation (SSH) had the highest stiffness followed by the construct with a stainless steel plate and locking screws (SSL) and were not statistically different from each other. Offset placement (SSLO) and using a titanium implant (TiL) significantly reduced construct stiffness. Fatigue Failure: The stainless steel with hybrid fixation group (SSH) withstood the most number of cycles to failure and higher loads, followed by the stainless steel plate and locking screw group (SSL), stainless steel plate with locking screws and offset group (SSLO), and the titanium plate and locking screws group (TiL) consecutively. Offset placement (SSLO) as well as using a titanium implant (TiL) reduced cycles to failure.
Conclusions: Using the same plate design, the study showed that implant material, screw type, and position of the plate affect the construct stiffness and fatigue life of the fixation construct. With this knowledge, the surgeon can decide the optimal construct based on a given fracture pattern, bone strength, and reduction quality.
Similar articles
-
The effect of surgeon-controlled variables on construct stiffness in lateral locked plating of distal femoral fractures.BMC Musculoskelet Disord. 2021 Jun 4;22(1):512. doi: 10.1186/s12891-021-04341-2. BMC Musculoskelet Disord. 2021. PMID: 34088275 Free PMC article.
-
Biomechanical Comparison of Volar Fixed-Angle Locking Plates for AO C3 Distal Radius Fractures: Titanium Versus Stainless Steel With Compression.J Hand Surg Am. 2015 Oct;40(10):2032-8. doi: 10.1016/j.jhsa.2015.06.098. Epub 2015 Aug 5. J Hand Surg Am. 2015. PMID: 26253601
-
Semi-rigid screws provide an auxiliary option to plate working length to control interfragmentary movement in locking plate fixation at the distal femur.Injury. 2015 Oct;46 Suppl 4:S24-32. doi: 10.1016/S0020-1383(15)30015-2. Injury. 2015. PMID: 26542863
-
[Biomechanical study of four palmar locking plates and one non-locking palmar plate for distal radius fractures: stiffness and load to failure tests in a cadaver model].Biomed Tech (Berl). 2009 Jun;54(3):150-8. doi: 10.1515/BMT.2009.018. Biomed Tech (Berl). 2009. PMID: 19469665 Review. German.
-
Biomechanical and biological aspects of defect treatment in fractures using helical plates.Acta Chir Orthop Traumatol Cech. 2014;81(4):267-71. Acta Chir Orthop Traumatol Cech. 2014. PMID: 25137496 Review.
Cited by
-
Micromotion-based balanced drilling technology to increase near cortical strain.BMC Surg. 2022 Nov 11;22(1):387. doi: 10.1186/s12893-022-01816-4. BMC Surg. 2022. PMID: 36368981 Free PMC article.
-
Opening-wedge osteotomies of the distal femur: minor advantages for a biplanar compared to a uniplanar technique.Knee Surg Sports Traumatol Arthrosc. 2019 Jul;27(7):2375-2384. doi: 10.1007/s00167-018-5332-5. Epub 2018 Dec 13. Knee Surg Sports Traumatol Arthrosc. 2019. PMID: 30547307
-
The Relation between the Radial Collapse and the Number of Metaphyseal Screws for Distal Radius Fractures.Malays Orthop J. 2022 Nov;16(3):30-35. doi: 10.5704/MOJ.2211.006. Malays Orthop J. 2022. PMID: 36589376 Free PMC article.
-
Finite Element Analysis of Fracture Fixation.Curr Osteoporos Rep. 2021 Aug;19(4):403-416. doi: 10.1007/s11914-021-00690-y. Epub 2021 Jun 29. Curr Osteoporos Rep. 2021. PMID: 34185266 Free PMC article. Review.
-
A systematic review of the use of titanium versus stainless steel implants for fracture fixation.OTA Int. 2021 Aug 18;4(3):e138. doi: 10.1097/OI9.0000000000000138. eCollection 2021 Sep. OTA Int. 2021. PMID: 34746670 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical