[Biomechanics of new implants for HTO]
- PMID: 28396901
- DOI: 10.1007/s00132-017-3417-3
[Biomechanics of new implants for HTO]
Abstract
Biomechanical characteristics of 5 tibial osteotomy plates for the treatment of medial knee joint osteoarthritis were examined. Fourth-generation tibial bone composites underwent a medial open-wedge high tibial osteotomy, using TomoFix™ standard, PEEKPower®, ContourLock®, TomoFix™ small stature plates, and iBalance® implants. Static compression load to failure and load-controlled cyclic fatigue failure tests were performed. All plates had sufficient stability up to 2400 N in the static compression load to failure tests. Screw breakage in the iBalance® group and opposite cortex fractures in all constructs occurred at lower loading conditions. The highest fatigue strength in terms of maximal load and number of cycles performed prior to failure was observed for the ContourLock® group followed by the iBalance® implants, the TomoFix™ standard and small stature plates. PEEKPower® had the lowest fatigue strength. All plates showed sufficient stability under static loading. Compared to the TomoFix™ and the PEEKPower® plates, the ContourLock® plate and iBalance® implant showed a higher mechanical fatigue strength during cyclic fatigue testing, suggesting that both mechanical static and fatigue strength increase with a wider proximal T‑shaped plate design together with diverging proximal screws. Mechanical strength of the bone-implant constructs decreases with a narrow T‑shaped proximal end design and converging proximal screws (TomoFix™) or a short vertical plate design (PEEKPower®). Published results indicate high fusion rates and good clinical results with the TomoFix™ plate, which is contrary to our findings. A certain amount of interfragmentary motion rather than high mechanical strength and stiffness seem to be important for bone healing which is outside the scope of this paper.
Keywords: Bone plates; High tibial osteotomy; Implant; Osteoarthritis; Tibial head.
Similar articles
-
Biomechanical properties of five different currently used implants for open-wedge high tibial osteotomy.J Exp Orthop. 2015 Dec;2(1):14. doi: 10.1186/s40634-015-0030-4. Epub 2015 Jun 18. J Exp Orthop. 2015. PMID: 26914882 Free PMC article.
-
Static and fatigue strength of a novel anatomically contoured implant compared to five current open-wedge high tibial osteotomy plates.J Exp Orthop. 2017 Dec 8;4(1):39. doi: 10.1186/s40634-017-0115-3. J Exp Orthop. 2017. PMID: 29222607 Free PMC article.
-
Stability of two angular stable locking plates for open wedge high tibial osteotomy (HTO): TomoFix™ versus LOQTEQ® HTO plate.Arch Orthop Trauma Surg. 2014 Oct;134(10):1437-42. doi: 10.1007/s00402-014-2042-4. Epub 2014 Jul 12. Arch Orthop Trauma Surg. 2014. PMID: 25015791
-
Biomechanical and biological aspects of defect treatment in fractures using helical plates.Acta Chir Orthop Traumatol Cech. 2014;81(4):267-71. Acta Chir Orthop Traumatol Cech. 2014. PMID: 25137496 Review.
-
High Tibial Osteotomy: Review of Techniques and Biomechanics.J Healthc Eng. 2019 May 2;2019:8363128. doi: 10.1155/2019/8363128. eCollection 2019. J Healthc Eng. 2019. PMID: 31191853 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources