Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Randomized Controlled Trial
. 2017 Jun 15;376(24):2319-2328.
doi: 10.1056/NEJMoa1614954. Epub 2017 Mar 29.

Bioresorbable Scaffolds versus Metallic Stents in Routine PCI

Collaborators, Affiliations
Free article
Randomized Controlled Trial

Bioresorbable Scaffolds versus Metallic Stents in Routine PCI

Joanna J Wykrzykowska et al. N Engl J Med. .
Free article

Abstract

Background: Bioresorbable vascular scaffolds were developed to overcome the shortcomings of drug-eluting stents in percutaneous coronary intervention (PCI). We performed an investigator-initiated, randomized trial to compare an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent in the context of routine clinical practice.

Methods: We randomly assigned 1845 patients undergoing PCI to receive either a bioresorbable vascular scaffold (924 patients) or a metallic stent (921 patients). The primary end point was target-vessel failure (a composite of cardiac death, target-vessel myocardial infarction, or target-vessel revascularization). The data and safety monitoring board recommended early reporting of the study results because of safety concerns. This report provides descriptive information on end-point events.

Results: The median follow-up was 707 days. Target-vessel failure occurred in 105 patients in the scaffold group and in 94 patients in the stent group (2-year cumulative event rates, 11.7% and 10.7%, respectively; hazard ratio, 1.12; 95% confidence interval [CI], 0.85 to 1.48; P=0.43); event rates were based on Kaplan-Meier estimates in time-to-event analyses. Cardiac death occurred in 18 patients in the scaffold group and in 23 patients in the stent group (2-year cumulative event rates, 2.0% and 2.7%, respectively), target-vessel myocardial infarction occurred in 48 patients in the scaffold group and in 30 patients in the stent group (2-year cumulative event rates, 5.5% and 3.2%), and target-vessel revascularization occurred in 76 patients in the scaffold group and in 65 patients in the stent group (2-year cumulative event rates, 8.7% and 7.5%). Definite or probable device thrombosis occurred in 31 patients in the scaffold group as compared with 8 patients in the stent group (2-year cumulative event rates, 3.5% vs. 0.9%; hazard ratio, 3.87; 95% CI, 1.78 to 8.42; P<0.001).

Conclusions: In this preliminary report of a trial involving patients undergoing PCI, there was no significant difference in the rate of target-vessel failure between the patients who received a bioresorbable scaffold and the patients who received a metallic stent. The bioresorbable scaffold was associated with a higher incidence of device thrombosis than the metallic stent through 2 years of follow-up. (Funded by Abbott Vascular; AIDA ClinicalTrials.gov number, NCT01858077 .).

PubMed Disclaimer

Comment in

Publication types

Associated data