Structural Shielding Design of a 6 MV Flattening Filter Free Linear Accelerator: Indian Scenario
- PMID: 28405104
- PMCID: PMC5370334
- DOI: 10.4103/jmp.JMP_99_16
Structural Shielding Design of a 6 MV Flattening Filter Free Linear Accelerator: Indian Scenario
Abstract
Detailed structural shielding of primary and secondary barriers for a 6 MV medical linear accelerator (LINAC) operated with flattening filter (FF) and flattening filter free (FFF) modes are calculated. The calculations have been carried out by two methods, one using the approach given in National Council on Radiation Protection (NCRP) Report No. 151 and the other based on the monitor units (MUs) delivered in clinical practice. Radiation survey of the installations was also carried out. NCRP approach suggests that the primary and secondary barrier thicknesses are higher by 24% and 26%. respectively, for a LINAC operated in FF mode to that of a LINAC operated in both FF and FFF modes with an assumption that only 20% of the workload is shared in FFF mode. Primary and secondary barrier thicknesses calculated from MUs delivered on clinical practice method also show the same trend and are higher by 20% and 19%, respectively, for a LINAC operated in FF mode to that of a LINAC operated in both FF and FFF modes. Overall, the barrier thickness for a LINAC operated in FF mode is higher about 20% to that of a LINAC operated in both FF and FFF modes.
Keywords: Flattening filter; flattening filter free; medical linear accelerator; monitor units; primary barrier; secondary barrier.
Conflict of interest statement
There are no conflicts of interest.
Figures
Similar articles
-
Impact of a flattening filter free linear accelerator on structural shielding design.Z Med Phys. 2014 Mar;24(1):38-48. doi: 10.1016/j.zemedi.2013.05.002. Epub 2013 Jun 10. Z Med Phys. 2014. PMID: 23763984
-
Beam Characterization of 10-MV Photon Beam from Medical Linear Accelerator without Flattening Filter.J Med Phys. 2017 Apr-Jun;42(2):65-71. doi: 10.4103/jmp.JMP_71_16. J Med Phys. 2017. PMID: 28706351 Free PMC article.
-
Treatment vault shielding for a flattening filter-free medical linear accelerator.Phys Med Biol. 2009 Mar 7;54(5):1265-73. doi: 10.1088/0031-9155/54/5/011. Epub 2009 Feb 3. Phys Med Biol. 2009. PMID: 19190359
-
Dosimetric parameters calculation for 18 MV photon beam in flattening filter (FF) and flattening filter free (FFF) linear accelerators with and without magnetic deflector and lead filter.Biomed Phys Eng Express. 2024 Feb 13;10(2). doi: 10.1088/2057-1976/ad2630. Biomed Phys Eng Express. 2024. PMID: 38316026
-
Current status and future perspective of flattening filter free photon beams.Med Phys. 2011 Mar;38(3):1280-93. doi: 10.1118/1.3554643. Med Phys. 2011. PMID: 21520840 Review.
Cited by
-
Comparison of Measured and Monte Carlo Calculated Dose Distributions from Indigenously Developed 6 MV Flattening Filter Free Medical Linear Accelerator.J Med Phys. 2018 Jul-Sep;43(3):162-167. doi: 10.4103/jmp.JMP_58_18. J Med Phys. 2018. PMID: 30305773 Free PMC article.
References
-
- Titt U, Vassiliev ON, Pönisch F, Dong L, Liu H, Mohan R. A flattening filter free photon treatment concept evaluation with Monte Carlo. Med Phys. 2006;33:1595–602. - PubMed
-
- Cashmore J. The characterization of unflattened photon beams from a 6 MV linear accelerator. Phys Med Biol. 2008;53:1933–46. - PubMed
-
- Georg D, Knöös T, McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys. 2011;38:1280–93. - PubMed
-
- Tsiamas P, Sajo E, Cifter F, Theodorou K, Kappas C, Makrigiorgos M, et al. Beam quality and dose perturbation of 6 MV flattening-filter-free linac. Phys Med. 2014;30:47–56. - PubMed
-
- Georg D, Kragl G, Wetterstedt Sa, McCavana P, McClean B, Knöös T. Photon beam quality variations of a flattening filter free linear accelerator. Med Phys. 2010;37:49–53. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources