Measurement considerations on examiner-dependent factors in the ultrasound assessment of developmental dysplasia of the hip

Affiliations

Measurement considerations on examiner-dependent factors in the ultrasound assessment of developmental dysplasia of the hip

Alexander Kolb et al. Int Orthop. 2017 Jun.

Abstract

Purpose: The standardized sonographic hip screening according to Graf has increased reliability and comparability of measurements in the screening of developmental dysplasia of the hip (DDH). However, examiner dependent factors have been discussed to influence sonographic measurements. The objectives of this study were to examine the tolerance of the transducer positioning and to analyse the impact of transducer inclinations on Graf's hip grading system.

Materials and methods: Twenty-four hips in consecutive newborns were screened sonographically in combination with an optoelectronic motion capture system to trace transducer positions in space. Subsequently five defined inclinations of the transducer relative to Graf's neutral transducer position were analysed, giving a total of 144 sonographic images.

Results: We found a permissible transducer inclination in the axial plane of 8.8° to anterior and 8.1° to posterior. In the frontal plane we found a permissible inclination of 15.4° to caudal and 7.2° to cranial. The impact on the α-angle was significant for posterior-cranial (p < 0.001), cranial (p = 0.009), and caudal (p < 0.001) inclined transducer positions. The effect on the results according to Graf's grading system was significant for the caudal inclination of the transducer position (p < 0.001).

Conclusion: Our findings show that the standardized plane defined by Graf's criteria allows notable inclinations of the transducer positions. Transducer inclinations show an impact on measurement results, which are clinically relevant. Those effects cannot be ruled out using Graf's ultrasound criteria alone. The examiner should pay attention to avoid transducer inclinations in the frontal plane and a combination of posterior and cranial inclination.

Keywords: Developmental Dysplasia of the Hip (DDH); Hip sonography.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Int Orthop. 2012 Jun;36(6):1243-6 - PubMed
    1. Orthopade. 1997 Jan;26(1):14-24 - PubMed
    1. Clin Orthop Relat Res. 1976 Sep;(119):6-10 - PubMed
    1. Swiss Med Wkly. 2004 Nov 13;134(45-46):671-7 - PubMed
    1. Pediatrics. 2010 Jan;125(1):e9-16 - PubMed

LinkOut - more resources