Dual-energy imaging method to improve the image quality and the accuracy of dose calculation for cone-beam computed tomography
- PMID: 28410679
- DOI: 10.1016/j.ejmp.2017.03.023
Dual-energy imaging method to improve the image quality and the accuracy of dose calculation for cone-beam computed tomography
Abstract
Purpose: To improve the image quality and accuracy of dose calculation for cone-beam computed tomography (CT) images through implementation of a dual-energy cone-beam computed tomography method (DE-CBCT), and evaluate the improvement quantitatively.
Methods: Two sets of CBCT projections were acquired using the X-ray volumetric imaging (XVI) system on a Synergy (Elekta, Stockholm, Sweden) system with 120kV (high) and 70kV (low) X-rays, respectively. Then, the electron density relative to water (relative electron density (RED)) of each voxel was calculated using a projection-based dual-energy decomposition method. As a comparison, single-energy cone-beam computed tomography (SE-CBCT) was used to calculate RED with the Hounsfield unit-RED calibration curve generated by a CIRS phantom scan with identical imaging parameters. The imaging dose was measured with a dosimetry phantom. The image quality was evaluated quantitatively using a Catphan 503 phantom with the evaluation indices of the reproducibility of the RED values, high-contrast resolution (MTF50%), uniformity, and signal-to-noise ratio (SNR). Dose calculation of two simulated volumetric-modulated arc therapy plans using an Eclipse treatment-planning system (Varian Medical Systems, Palo Alto, CA, USA) was performed on an Alderson Rando Head and Neck (H&N) phantom and a Pelvis phantom. Fan-beam planning CT images for the H&N and Pelvis phantom were set as the reference. A global three-dimensional gamma analysis was used to compare dose distributions with the reference. The average gamma values for targets and OAR were analyzed with paired t-tests between DE-CBCT and SE-CBCT.
Results: In two scans (H&N scan and body scan), the imaging dose of DE-CBCT increased by 1.0% and decreased by 1.3%. It had a better reproducibility of the RED values (mean bias: 0.03 and 0.07) compared with SE-CBCT (mean bias: 0.13 and 0.16). It also improved the image uniformity (57.5% and 30.1%) and SNR (9.7% and 2.3%), but did not affect the MTF50%. Gamma analyses of the 3D dose distribution with criteria of 1%/1mm showed a pass rate of 99.0-100% and 85.3-97.6% for DE-CBCT and 73.5-99.1% and 80.4-92.7% for SE-CBCT. The average gamma values were reduced significantly by DE-CBCT (p< 0.05). Gamma index maps showed that matching of the dose distribution between CBCT-based and reference was improved by DE-CBCT.
Conclusions: DE-CBCT can achieve both better image quality and higher accuracy of dose calculation, and could be applied to adaptive radiotherapy.
Keywords: Adaptive radiotherapy; CBCT; Dose calculation; Dual-energy imaging; Image quality.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Dose calculation accuracy of different image value to density tables for cone-beam CT planning in head & neck and pelvic localizations.Phys Med. 2015 Mar;31(2):146-51. doi: 10.1016/j.ejmp.2014.12.007. Epub 2015 Jan 13. Phys Med. 2015. PMID: 25595131
-
Assessment of image quality and dose calculation accuracy on kV CBCT, MV CBCT, and MV CT images for urgent palliative radiotherapy treatments.J Appl Clin Med Phys. 2016 Mar 8;17(2):279-290. doi: 10.1120/jacmp.v17i2.6040. J Appl Clin Med Phys. 2016. PMID: 27074487 Free PMC article.
-
Improved image quality of cone beam CT scans for radiotherapy image guidance using fiber-interspaced antiscatter grid.Med Phys. 2014 Jun;41(6):061910. doi: 10.1118/1.4875978. Med Phys. 2014. PMID: 24877821
-
Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography.J Cancer Res Ther. 2019 Jul-Sep;15(3):470-474. doi: 10.4103/jcrt.JCRT_949_17. J Cancer Res Ther. 2019. PMID: 31169206 Review.
-
[External beam radiotherapy cone beam-computed tomography-based dose calculation].Cancer Radiother. 2018 Feb;22(1):85-100. doi: 10.1016/j.canrad.2017.07.050. Epub 2017 Dec 21. Cancer Radiother. 2018. PMID: 29276135 Review. French.
Cited by
-
Fast-switching dual energy cone beam computed tomography using the on-board imager of a commercial linear accelerator.Phys Med Biol. 2020 Jan 13;65(1):015013. doi: 10.1088/1361-6560/ab5c35. Phys Med Biol. 2020. PMID: 31775131 Free PMC article.
-
Online daily assessment of dose change in head and neck radiotherapy without dose-recalculation.J Appl Clin Med Phys. 2018 Sep;19(5):659-665. doi: 10.1002/acm2.12432. Epub 2018 Aug 7. J Appl Clin Med Phys. 2018. PMID: 30084159 Free PMC article.
-
Effect of computed tomography value error on dose calculation in adaptive radiotherapy with Elekta X-ray volume imaging cone beam computed tomography.J Appl Clin Med Phys. 2021 Sep;22(9):271-279. doi: 10.1002/acm2.13384. Epub 2021 Aug 10. J Appl Clin Med Phys. 2021. PMID: 34375008 Free PMC article.
-
Optimization-based image reconstruction regularized with inter-spectral structural similarity for limited-angle dual-energy cone-beam CT.Phys Med Biol. 2025 Jul 11;70(14):145010. doi: 10.1088/1361-6560/ade843. Phys Med Biol. 2025. PMID: 40562072 Free PMC article.
-
Trends in the Use of Weightbearing Computed Tomography.J Clin Med. 2024 Sep 18;13(18):5519. doi: 10.3390/jcm13185519. J Clin Med. 2024. PMID: 39337007 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous