. 2017 Jun;50(9):513-520.
doi: 10.1016/j.clinbiochem.2017.04.006. Epub 2017 Apr 12.

Effects of preanalytical factors on hemoglobin measurement: A comparison of two HemoCue® point-of-care analyzers

Affiliations
Clinical Trial

Effects of preanalytical factors on hemoglobin measurement: A comparison of two HemoCue® point-of-care analyzers

Ralph D Whitehead Jr et al. Clin Biochem. 2017 Jun.

Abstract

Background: In field studies, hemoglobin (Hb) is often measured using a battery-operated, portable HemoCue® hemoglobinometer.

Methods: We compared the performance of 2 HemoCue® models (Hb-201+ and Hb-301) and investigated effects of preanalytical factors on Hb results by simulating unfavorable field conditions.

Results: The Hb-301 produced 2.6% higher results compared to the Hb-201+. Hb had to be measured within 1min of filling the Hb-301 cuvette to avoid artificially elevated concentrations (1.3% per min). The Hb-301 cuvettes withstood elevated temperature (37°C) and humidity (72%) for 3weeks, while the Hb-201+ cuvettes degraded within 10min under those conditions. Both cuvette types withstood elevated temperature for 3weeks. Properly-collected venous and capillary blood produced comparable results. Pooled capillary blood produced comparable results to the second and third but not the fourth drop of blood (3.3% lower). Blood could be stored for ≤4days at 10-30°C before Hb-201+ measurement, but only for 1day at 10-23°C before Hb-301 measurement (≤1% change in Hb).

Conclusions: Higher Hb results obtained with the Hb-301 may influence the interpretation of anemia prevalence in health surveys. While the Hb-301 performed better in high humidity conditions, the Hb-201+ provided more user flexibility regarding delayed Hb reading.

Keywords: Capillary blood; Elevated temperature; Freeze-thawing; Humidity; Storage stability; Venous blood.

PubMed Disclaimer

References

    1. World Health Organization (WHO) Worldwide prevalence of anemia, 1993–2005. In: de Benoist Bruno, McLean Erin, Egli Ines, Cogswell Mary., editors. WHO global database on anemia. Geneva: World Health Organization; 2008. Available from: http://whqlibdoc.who.int/publications/2008/9789241596657_eng.pdf?ua=1.
    1. Northrop-Clewes CA, Thurnham DI. Biomarkers for the differentiation of anemia and their clinical usefulness. J Blood Med. 2013;4:11–22. - PMC - PubMed
    1. Sanchis-Gomar F, Cortell-Ballester J, Pareja-Galeano H, Banfi G, Lippi G. Hemoglobin Point-of-Care testing: the HemoCue system. J Lab Autom. 2013;18:198–205. - PubMed
    1. Jaggernath M, Naicker R, Madurai S, Brockman MA, Ndung’u T, Gelderblom HC. Diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 Point-of-Care hemoglobin meters in a central laboratory and a community based clinic in Durban, South Africa. PLOS ONE. 2016;11(4):e0152184. doi: 10.1371/journal.pone.0152184. - DOI - PMC - PubMed
    1. Monárrez-Espino J. Comparison of the analytic performance between the B-HB and HB-201+ HemoCue® hemoglobinometers for venous and capillary blood under field work conditions. Ecol Food Nutr. 2008;47:159–69.