3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times
- PMID: 28432504
- DOI: 10.1007/s00330-017-4816-x
3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times
Abstract
Objectives: To investigate optimised isotropic 3D turbo spin echo (TSE) and gradient echo (GRE)-based pulse sequences for visualisation of articular cartilage lesions within the knee joint.
Methods: Optimisation of experimental imaging sequences was completed using healthy volunteers (n=16) with a 3-Tesla (3T) MRI scanner. Imaging of patients with knee cartilage abnormalities (n=57) was then performed. Acquired sequences included 3D proton density-weighted (PDW) TSE (SPACE) with and without fat-suppression (FS), and T2*W GRE (TrueFISP) sequences, with acquisition times of 6:51, 6:32 and 5:35 min, respectively.
Results: One hundred sixty-one confirmed cartilage lesions were detected and categorised (Grade II n=90, Grade III n=71). The highest sensitivity and specificity for detecting cartilage lesions were obtained with TrueFISP with values of 84.7% and 92%, respectively. Cartilage SNR mean for PDW SPACE-FS was the highest at 72.2. TrueFISP attained the highest CNR means for joint fluid/cartilage (101.5) and joint fluid/ligament (156.5), and the lowest CNR for cartilage/meniscus (48.5). Significant differences were identified across the three sequences for all anatomical structures with respect to SNR and CNR findings (p-value <0.05).
Conclusion: Isotropic TrueFISP at 3T, optimised for acquisition time, accurately detects cartilage defects, although it demonstrated the lowest contrast between cartilage and meniscus.
Key points: • Cartilage is better visualised with 3D TrueFISP than 3D SPACE sequences. • 3D TrueFISP is a reliable sequence for detecting low- and high-grade cartilage defects. • 3D TrueFISP at 3T provides excellent contrast between cartilage and joint fluid.
Keywords: Articular; Cartilage; Evidence-based practice; Knee joint; Magnetic resonance imaging.
Similar articles
-
Evaluation of optimised 3D turbo spin echo and gradient echo MR pulse sequences of the knee at 3T and 1.5T.Radiography (Lond). 2021 May;27(2):389-397. doi: 10.1016/j.radi.2020.09.020. Epub 2020 Oct 7. Radiography (Lond). 2021. PMID: 33036913
-
MRI of the knee at 3T: first clinical results with an isotropic PDfs-weighted 3D-TSE-sequence.Invest Radiol. 2009 Sep;44(9):585-97. doi: 10.1097/RLI.0b013e3181b4c1a1. Invest Radiol. 2009. PMID: 19668001
-
Three-dimensional magnetic resonance observation of cartilage repair tissue (MOCART) score assessed with an isotropic three-dimensional true fast imaging with steady-state precession sequence at 3.0 Tesla.Invest Radiol. 2009 Sep;44(9):603-12. doi: 10.1097/RLI.0b013e3181b5333c. Invest Radiol. 2009. PMID: 19692843 Clinical Trial.
-
Advanced morphological and biochemical magnetic resonance imaging of cartilage repair procedures in the knee joint at 3 Tesla.Semin Musculoskelet Radiol. 2008 Sep;12(3):196-211. doi: 10.1055/s-0028-1083104. Epub 2008 Oct 10. Semin Musculoskelet Radiol. 2008. PMID: 18850502 Review.
-
2D versus 3D MRI of osteoarthritis in clinical practice and research.Skeletal Radiol. 2023 Nov;52(11):2211-2224. doi: 10.1007/s00256-023-04309-4. Epub 2023 Mar 13. Skeletal Radiol. 2023. PMID: 36907953 Review.
Cited by
-
Patterns of cartilage degeneration in knees with medial tibiofemoral offset.Skeletal Radiol. 2019 Jun;48(6):931-937. doi: 10.1007/s00256-018-3093-3. Epub 2018 Oct 24. Skeletal Radiol. 2019. PMID: 30357459
-
Comparison of 3D Gradient-Echo Versus 2D Sequences for Assessing Shoulder Joint Image Quality in MRI.Int J Biomed Imaging. 2024 Oct 11;2024:2244875. doi: 10.1155/2024/2244875. eCollection 2024. Int J Biomed Imaging. 2024. PMID: 39429699 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical