Subacute right ventricular pacemaker lead perforation: evaluation by echocardiography and cardiac CT

Affiliations
Case Reports

Subacute right ventricular pacemaker lead perforation: evaluation by echocardiography and cardiac CT

Reinier P J Boxma et al. J Echocardiogr. 2017 Dec.
No abstract available

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
a Chest radiography [postero-anterior (PA) view] the day after pacemaker implantation. b Lateral view. c Chest radiography (PA view) upon presentation to the emergency room showing an altered position of the RV lead (arrowhead). d Lateral view (arrowhead shows altered lead position). e Transthoracic echocardiography (subcostal view) suggesting myocardial perforation of the RV lead through the RV apex (arrowhead), but without pericardial effusion, raising doubt as to whether the lead tip lay in the pericardial space. See also online video. LA left atrium, LV left ventricle, RA right atrium, RV right ventricle. f Thoracic computed tomography (coronal view) with lead tip clearly visible through the myocardium (arrowhead). RV right ventricle. g Thoracic computed tomography (sagittal view) showing lead tip running through myocardium (arrowhead). RV right ventricle

References

    1. Ellenbogen KA, Wood MA, Shepard RK. Delayed complications following pacemaker implantation. Pacing Clin Electrophysiol. 2002;25:1155–1158. doi: 10.1046/j.1460-9592.2002.01155.x. - DOI - PubMed
    1. Cano Ó, Andrés A, Alonso P, Osca J, Sancho-Tello MJ, Olagüe J, Martínez-Dolz L. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads. Europace. 2017;19:96–102. - PubMed
    1. Sterlinski M, Przybylski A, Maciag A, Syska P, Pytkowski M, Lewandowski M, Kowalik I, Firek B, Kołsut P, Religa G, Kuśmierczyk M, Walczak F, Szwed H. Subacute cardiac perforations associated with active fixation leads. Europace. 2009;11:206–212. doi: 10.1093/europace/eun363. - DOI - PubMed
    1. Rajkumar CA, Claridge S, Jackson T, Behar J, Johnson J, Sohal M, Amraoui S, Nair A, Preston R, Gill J, Rajani R, Aldo Rinaldi C. Diagnosis and management of iatrogenic cardiac perforation caused by pacemaker and defibrillator leads. Europace. 2016 - PubMed
    1. Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007;30:28–32. doi: 10.1111/j.1540-8159.2007.00575.x. - DOI - PubMed