Effects of Propofol on Respiratory Drive and Patient-ventilator Synchrony during Pressure Support Ventilation in Postoperative Patients: A Prospective Study
- PMID: 28485314
- PMCID: PMC5443020
- DOI: 10.4103/0366-6999.205864
Effects of Propofol on Respiratory Drive and Patient-ventilator Synchrony during Pressure Support Ventilation in Postoperative Patients: A Prospective Study
Abstract
Background: Propofol is increasingly used during partial support mechanical ventilation such as pressure support ventilation (PSV) in postoperative patients. However, breathing pattern, respiratory drive, and patient-ventilator synchrony are affected by the sedative used and the sedation depth. The present study aimed to evaluate the physiologic effects of varying depths of propofol sedation on respiratory drive and patient-ventilator synchrony during PSV in postoperative patients.
Methods: Eight postoperative patients receiving PSV for <24 h were enrolled. Propofol was administered to achieve and maintain a Ramsay score of 4, and the inspiratory pressure support was titrated to obtain a tidal volume (VT) of 6-8 ml/kg. Then, the propofol dose was reduced to achieve and maintain a Ramsay score of 3 and then 2. At each Ramsay level, the patient underwent 30-min trials of PSV. We measured the electrical activity of the diaphragm, flow, airway pressure, neuro-ventilatory efficiency (NVE), and patient-ventilator synchrony.
Results: Increasing the depth of sedation reduced the peak and mean electrical activity of the diaphragm, which suggested a decrease in respiratory drive, while VT remained unchanged. The NVE increased with an increase in the depth of sedation. Minute ventilation and inspiratory duty cycle decreased with an increase in the depth of sedation, but this only achieved statistical significance between Ramsay 2 and both Ramsay 4 and 3 (P < 0.05). The ineffective triggering index increased with increasing sedation depth (9.5 ± 4.0%, 6.7 ± 2.0%, and 4.2 ± 2.1% for Ramsay 4, 3, and 2, respectively) and achieved statistical significance between each pair of depth of sedation (P < 0.05). The depth of sedation did not affect gas exchange.
Conclusions: Propofol inhibits respiratory drive and deteriorates patient-ventilator synchrony to the extent that varies with the depth of sedation. Propofol has less effect on breathing pattern and has no effect on VT and gas exchange in postoperative patients with PSV.
Conflict of interest statement
There are no conflicts of interest.
Figures
References
-
- Younes M, Brochard L, Grasso S, Kun J, Mancebo J, Ranieri M, et al. A method for monitoring and improving patient: Ventilator interaction. Intensive Care Med. 2007;33:1337–46. doi: 10.1007/s00134-007-0681-4. - PubMed
-
- Colombo D, Cammarota G, Bergamaschi V, De Lucia M, Corte FD, Navalesi P. Physiologic response to varying levels of pressure support and neurally adjusted ventilatory assist in patients with acute respiratory failure. Intensive Care Med. 2008;34:2010–8. doi: 10.1007/s00134-008-1208-3. - PubMed
-
- Thille AW, Rodriguez P, Cabello B, Lellouche F, Brochard L. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2006;32:1515–22. doi: 10.1007/s00134-006-0301-8. - PubMed
-
- Tobin MJ, Jubran A, Laghi F. Patient-ventilator interaction. Am J Respir Crit Care Med. 2001;163:1059–63. doi: 10.1164/ajrccm.163.5.2005125. - PubMed
-
- Appendini L, Purro A, Patessio A, Zanaboni S, Carone M, Spada E, et al. Partitioning of inspiratory muscle workload and pressure assistance in ventilator-dependent COPD patients. Am J Respir Crit Care Med. 1996;154:1301–9. doi: 10.1164/ajrccm.154.5.8912740. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources