Is it acceptable to use coagulation plasma samples stored at room temperature and 4°C for 24 hours for additional prothrombin time, activated partial thromboplastin time, fibrinogen, antithrombin, and D-dimer testing?
- PMID: 28488785
- DOI: 10.1111/ijlh.12664
Is it acceptable to use coagulation plasma samples stored at room temperature and 4°C for 24 hours for additional prothrombin time, activated partial thromboplastin time, fibrinogen, antithrombin, and D-dimer testing?
Abstract
Introduction: Coagulation laboratories are faced on daily basis with requests for additional testing in already analyzed fresh plasma samples. This prompted us to examine whether plasma samples stored at room temperature (RT), and 4°C for 24 hours can be accepted for additional prothrombin time (PT), activated partial thromboplastin time (aPTT), fibrinogen (Fbg), antithrombin (AT), and D-dimer testing.
Methods: We measured PT, aPTT, Fbg in 50 and AT in 30 plasma samples with normal and pathological values, within 4 hours of blood collection (baseline results) and after 24-hours storage at RT (primary tubes), and 4°C (aliquots). D-dimer stability was investigated in 20 samples stored in primary tubes at 4°C.
Results: No statistically significant difference between baseline results and results in samples stored at RT and 4°C was observed for PT (P=.938), aPTT (P=.186), Fbg (P=.962), AT (P=.713), and D-dimers (P=.169). The highest median percentage changes were found for aPTT, being more pronounced for samples stored at 4°C (13.0%) than at RT (8.7%).
Conclusion: Plasma samples stored both at RT and 4°C for 24 hours are acceptable for additional PT, Fbg, and AT testing. Plasma samples stored 24 hours in primary tubes at 4°C are suitable for D-dimer testing.
Keywords: 24-hours plasma stability; D-dimer; antithrombin; routine coagulation tests; storage temperature.
© 2017 John Wiley & Sons Ltd.
Similar articles
-
Effect of freezing plasma at -20°C for 2 weeks on prothrombin time, activated partial thromboplastin time, dilute Russell viper venom time, activated protein C resistance, and D-dimer levels.Clin Appl Thromb Hemost. 2015 Jan;21(1):41-7. doi: 10.1177/1076029614535973. Epub 2014 May 19. Clin Appl Thromb Hemost. 2015. PMID: 24842886
-
Influence of 8 and 24-h storage of whole blood at ambient temperature on prothrombin time, activated partial thromboplastin time, fibrinogen, thrombin time, antithrombin and D-dimer.Blood Coagul Fibrinolysis. 2011 Apr;22(3):215-20. doi: 10.1097/MBC.0b013e328343f8bf. Blood Coagul Fibrinolysis. 2011. PMID: 21297453
-
Impact of different storage times at room temperature of unspun citrated blood samples on routine coagulation tests results. Results of a bicenter study and review of the literature.Int J Lab Hematol. 2017 Oct;39(5):458-468. doi: 10.1111/ijlh.12660. Epub 2017 May 8. Int J Lab Hematol. 2017. PMID: 28481034
-
Routine Coagulation.Clin Lab Med. 2024 Sep;44(3):527-539. doi: 10.1016/j.cll.2024.04.012. Epub 2024 Jun 4. Clin Lab Med. 2024. PMID: 39089756 Review.
-
Integrating biomarkers for hemostatic disorders into computational models of blood clot formation: A systematic review.Math Biosci Eng. 2024 Dec 4;21(12):7707-7739. doi: 10.3934/mbe.2024339. Math Biosci Eng. 2024. PMID: 39807050
Cited by
-
Diagnostic value of D-dimer for lower extremity deep venous thrombosis caused by rib fracture: a retrospective study.J Orthop Surg Res. 2023 Jul 20;18(1):515. doi: 10.1186/s13018-023-03997-x. J Orthop Surg Res. 2023. PMID: 37475021 Free PMC article.
-
In vitro assessment of quality of citrate-phosphate-dextrose-adenine-1 preserved feline blood collected by a commercial closed system.J Vet Intern Med. 2018 May;32(3):1051-1059. doi: 10.1111/jvim.15056. Epub 2018 Apr 10. J Vet Intern Med. 2018. PMID: 29635743 Free PMC article.
-
Micro-mechanical blood clot testing using smartphones.Nat Commun. 2022 Feb 11;13(1):831. doi: 10.1038/s41467-022-28499-y. Nat Commun. 2022. PMID: 35149711 Free PMC article.